Settings conflict;
-221
burst count reduced
Since the burst period is currently at its maximum, the function generator
has reduced the burst count to allow the specified waveform frequency.
-221
Settings conflict;
trigger delay reduced to fit entire burst
The function generator has reduced the trigger delay in order to maintain
the current burst period and burst count. The trigger delay is the time
between the receipt of the trigger and the start of the burst waveform.
-221
Settings conflict;
triggered burst not available for noise
You cannot use the noise function in the triggered burst mode. Noise is
allowed only in the gated burst mode.
Settings conflict;
-221
amplitude units changed to Vpp due to high-Z load
The output units (VOLT:UNIT command) cannot be set to dBm if the
output termination is currently set to "high impedance" (OUTP:LOAD
command). The function generator has converted the units to Vpp.
-221
Settings conflict;
trigger output disabled by trigger external
When the external trigger source is selected (TRIG:SOUR EXT command),
the function generator automatically disables the "trigger out" signal.
The rear-panel Trig connector cannot be used for both operations at the
same time (an externally-triggered waveform uses the same connector to
trigger the sweep or burst).
-221
Settings conflict;
trigger output connector used by FSK
If you have enabled FSK and have selected the external trigger source
(FSK:SOUR EXT command), the "trigger out" signal cannot be enabled
(OUTP:TRIG ON command). The rear-panel Trig connector cannot be
used for both operations at the same time.
Chapter 5 Error Messages
Execution Errors
4
5
253