3
Chapter 3 Features and Functions
Pulse Waveforms

Edge Time

The edge time represents the time from the 10% threshold to the 90%
threshold of both the rising and falling edges.
• Edge time: 5 ns to 1 ms (see restrictions below). The default edge time
is 5 ns.
• The specified edge time must fit within the specified pulse width as
shown below. The function generator will adjust the edge time as
needed to accommodate the specified pulse width.
Edge Time < 0.625 X Pulse Width
• Front-Panel Operation: After selecting the pulse function, press the
Edge Time softkey. Then use the knob or numeric keypad to enter the
desired edge time.
• Remote Interface Operation:
PULSe:TRANsition {<seconds>|MINimum|MAXimum}