Chapter 5 Error Messages

Execution Errors

Execution Errors
Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was
ignored. Make sure that you have selected the proper trigger source and
verify that the sweep or burst mode is enabled.
Too much data
An arbitrary waveform was specified that contains more than 65,536
waveform points. Verify the number of points in the DATA VOLATILE or
DATA:DAC VOLATILE command.
Settings conflict;
turned off infinite burst to allow immediate trigger source
An infinite count burst is allowed only when the external or bus
(software) trigger source is selected. The burst count has been set to the
maximum value (1,000,000 cycles).
Settings conflict;
infinite burst changed trigger source to BUS
An infinite count burst is allowed only when the external or bus
(software) trigger source is selected. Sending the BURS:NCYC INF
command has automatically changed the trigger source from immediate
to bus.
Settings conflict;
burst period increased to fit entire burst
The number of cycles specified by the BURS:NCYC command takes
priority over the burst period (as long as the burst period is not at its
maximum value). The function generator has increased the burst period
to accommodate the specified burst count or waveform frequency.