Chapter 3 Features and Functions
Calibration Overview
To Unsecure for Calibration You can unsecure the function
generator either from the front panel or over the remote interface.
The function generator is secured when shipped from the factory and
the security code is set to "AT33250A".
• Once you enter a security code, that code must be used for both front-
panel and remote operation. For example, if you secure the function
generator from the front panel, you must use that same code to
unsecure it from the remote interface.
• Front-Panel Operation: Press
softkey from the "Test / Cal" menu.
• Remote Interface Operation: To unsecure the function generator,
send the following command with the correct security code.
CAL:SECURE:STATE OFF,AT33250A
To Secure Against Calibration You can secure the function generator
either from the front panel or over the remote interface. The function
generator is secured when shipped from the factory and the security code
is set to "AT33250A".
• Once you enter a security code, that code must be used for both front-
panel and remote operation. For example, if you secure the function
generator from the front panel, you must use that same code to
unsecure it from the remote interface. .
• Front-Panel Operation: Press
softkey from the "Test / Cal" menu.
• Remote Interface Operation: To secure the function generator,
send the following command with the correct security code.
CAL:SECURE:STATE ON,AT33250A