The Standard Event Register

The Standard Event register group reports the following types of events:
power-on detected, command syntax errors, command execution errors,
self-test or calibration errors, query errors, or the *OPC command has
been executed. Any or all of these conditions can be reported to the
Standard Event summary bit through the enable register. To set the
enable register mask, you must write a decimal value to the register
using the *ESE command.
Bit Definitions – Standard Event Register
Bit Number
0 Operation Complete
1 Not Used
2 Query Error
3 Device Error
4 Execution Error
5 Command Error
6 Not Used
7 Power On
Chapter 4 Remote Interface Reference
Decimal
Value
Definition
1
All commands prior to and including *OPC
have completed and the overlapped command
(e.g., *TRG for burst) has completed.
2
Not Used. Returns "0".
4
The instrument tried to read the output buffer
but it was empty. Or, a new command line
was received before a previous query has
been read. Or, both the input and output
buffers are full.
8
A self-test, cal, or other device-specific error
has occurred (see chapter 5).
16
An execution error has occurred
(see chapter 5).
32
A command syntax error has occurred
(see chapter 5).
64
Not Used. Returns "0".
128
Power has been cycled on since the last time
the event register was read or cleared.
The SCPI Status System
233