Repair and Replacement Procedures

Removing and Replacing the A27 and A28 Mixer Bricks

Removing and Replacing the A27 and A28 Mixer Bricks
Tools Required
• T-8 TORX driver (set to 6 in-lb or 0.68N.m)
• T-10 TORX driver (set to 9 in-lb or 1.02 N.m)
• T-20 TORX driver (set to 21 in-lb or 2.38 N.m)
• 5/16 inch open-end torque wrench (set to 1.13 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-9
for this procedure.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of the cables
is necessary to remove the assembly. However, do not over-bend the semirigid cables.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
3. Position the analyzer bottom side up.
4. Remove the A26 splitter. Refer to
5. The mixer bricks and the mounting blocks to which they are attached, must be removed as a complete
assembly. 
Disconnect all semirigid cables (item
6. Disconnect the ribbon cables (item
7. Remove ten screws (item
8. Move the disconnected semirigid cables out of the way and lift the mixer brick mounting block out of the
analyzer just enough to allow the gray flexible RF cables (item
necessary to loosen the connector at the other end of some of the semirigid cables to allow them to be
moved enough.
Note the locations of each of the gray flexible cables for reconnection later. Disconnect these cables
and remove the mixer brick mounting block with the mixer brick(s) from the analyzer.
9. If replacing a mixer brick, remove 3 screws (not shown) that fasten each mixer brick to the mixer brick
mounting block.
10. If replacing the A28 mixer brick, remove the 3 dB pads (item
replacement mixer brick.
7-30
"Removing the Covers" on page
"Removing and Replacing the A26 Splitter" on page
) from the mixer bricks on the mounting blocks to be removed.
) from each of the mixer bricks.
) from the mixer brick mounting block.