Removing And Replacing The A42-A45 Bias Tees - Keysight N5227A Service Manual

2-port and 4-port pna microwave network analyzer 10 mhz - 67 ghz
Table of Contents

Advertisement

Repair and Replacement Procedures
Removing and Replacing the A42–A45 Bias Tees
Removing and Replacing the A42–A45 Bias Tees
Tools Required
• T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
• T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
• 5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-16
for this procedure.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of the cables
is necessary to remove the assembly. However, do not over-bend the semirigid cables.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
3. Position the analyzer bottom side up.
4. The bias tees and the blocks to which they are attached, must be removed as a complete assembly.
Disconnect four semirigid cables (item
5. Disconnect the wire harness cable (item
6. Remove two test set deck screws (item
7. Move the disconnected semirigid cables out of the way and remove the bias tee block from the analyzer.
It may be necessary to loosen the connector at the other end of some of the semirigid cables to allow
them to be moved enough.
8. Remove two screws from the bias tee in order to remove it from the bias tee block.
Replacement Procedure
1. Reverse the order of the removal procedure.
Orient the position of the bias tees and block as shown in
Torque all cable connections to 10 in-lbs.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to
Table 7-2 on page
7-40
"Removing the Covers" on page
) from the bias tee block to be removed.
) from the bias tee.
) from the bias tee block to be removed.
Figure
7-54.
PNA Series Microwave Network Analyzers
7-6.
7-17.
Service Guide N5227-90001
N5227A