Keysight N5227A Service Manual page 59

2-port and 4-port pna microwave network analyzer 10 mhz - 67 ghz
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5227A
If the Analyzer Fails this Test:
• Run the Receiver Characterization adjustment, and repeat this test.
• If the analyzer still fails this test, replace the A23 mixer brick for a Port 1 or Port 2 failure or the A24 mixer
brick for a Port 3 or Port 4 failure, then repeat this test. Refer to
A28 Mixer Bricks" on page
Service Guide N5227-90001
7-30.
Tests and Adjustments
Performance Tests
"Removing and Replacing the A27 and
3-35

Advertisement