Removing the Covers

Removing the Covers
Tools Required
• T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
• T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
Refer to
Figure 6-38, "External Hardware and Miscellaneous Parts, All Options,"
Removing the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front handles in the vertical
position. Do not place the analyzer on its front panel without the handles. This will damage
the front panel assemblies.
1. Disconnect the power cord.
2. Remove the strap handles (item
ends until the handle is free of the analyzer.
3. Remove the foot locks (item
feet from the outer cover.
4. Remove the four rear panel feet (item
driver.
5. Slide the outer cover toward the rear of the analyzer and remove it.
Removing the Inner Cover
1. With a T-10 TORX driver, remove the pan head screws (item
2. With a T-10 TORX driver, remove the flat head screws (item
3. Lift off the cover (item
Replacement Procedure
1. On the top side of the PNA, carefully position the grey flex cables so they can't be pinched between the
covers and the rails.
2. On the bottom side of the PNA, carefully fold or push down the ribbon cables and wires so they can't be
pinched between the hardware and the outer cover. Ribbon cables and wires must never be positioned
on top of hardware.
3. Reverse the order of the removal procedures above.
7-6
) by loosening the screws (item
) from the four bottom feet (item
) by removing the center screws (item
).
PNA Series Microwave Network Analyzers
for these procedures.
), with a T-20 TORX driver, on both
) and then remove the four bottom
) with a T-20 TORX
).
).
Service Guide N5227-90001
N5227A