Keysight N5227A Service Manual page 111

2-port and 4-port pna microwave network analyzer 10 mhz - 67 ghz
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5227A
Checking the A27 and A28 Mixer Brick Receiver Outputs
Set the network analyzer for an S
1. Locate the following flexible cables at the receiver IF outputs of the mixer brick(s).
• 2-port models
— Receiver A; W61 of A27
— Receiver R1; W62 of A27
— Receiver R2; W63 of A27
— Receiver B; W64 of A27
• 4-port models
— Receiver A; W61 of A27
— Receiver R1; W62 of A27
— Receiver R2; W63 of A27
— Receiver B; W64 of A27
— Receiver C; W68 of A28
— Receiver R3;
W67
— Receiver R4; W66 of A28
— Receiver D; W65 of A28
2. Disconnect the flexible cable at the suspect receiver.
3. Connect the spectrum analyzer to the suspect receiver connector.
4. Refer to
the IMPORTANT notice on page
7.61 MHz (analyzers with DSP version 4.0), or 7.44 MHz (analyzers with DSP version 5.0), and 1 GHz.
• If the measured signal is present, continue testing at
page
4-42.
5. If the measured signal is missing on the R1 receiver, continue testing at
Mixer Switch" on page
If the measured signal is missing on any receiver (other than the R1 receiver noted above) and the
analyzer does not have receiver attenuators, replace the A27 or A28 mixer brick, whichever is
appropriate. Refer to
If the measured signal is missing on the A, B, C, or D receivers and the analyzer has receiver attenuators,
continue testing at
"Checking the A46, A47, A48, and A49 35-dB Receiver Step Attenuators" on
page
4-42.
Checking the A37 Reference Mixer Switch
1. Remove the front panel REF 1 jumper and connect a spectrum analyzer to the front-panel REF 1 SOURCE
OUT connector.
2. If the measured signal is present, replace the A27 mixer brick. Refer to
and A28 Mixer Bricks" on page
Service Guide N5227-90001
measurement with a CW frequency of 1 GHz.
11
of A28
4-23. The measured signal on the spectrum analyzer should be at
4-41.
"Removing and Replacing the A27 and A28 Mixer Bricks" on page
7-30.
Measurement System Troubleshooting
"Checking the A24 IF Multiplexer Board" on
"Checking the A37 Reference
"Removing and Replacing the A27
Troubleshooting
7-30.
4-41