General Purpose Maintenance Procedures
Option Enable Utility

Enabling or Removing Options

There are two types of options:
• Hardware: Hardware options involve adding additional hardware to the analyzer. After the proper
hardware has been installed in the analyzer, the option can be enabled using the option enable utility.
It is necessary to re-enable all installed hardware options if the test set motherboard is replaced.
• Software: Software options add features or functionality to the analyzer without the need for additional
hardware. These options are enabled using the option enable utility and require a special license key.
NOTE
Some applications require a license key that is provided by Keysight. If you do not have the
required license key, contact Keysight for assistance. Refer to
page
2-6.
It is necessary to backup all installed software options if the frequency reference board is replaced.
To enable or remove an option:
1. Start the option enable utility. Refer to
2. Click the arrow in the
appear.
020 - Add IF inputs
021 - Pulse Modulator on Source 1
022 - Pulse Modulator on Source 2
025 - 4-Receiver Pulse Generator
086 - Gain Compression Application
087- IMD Measurements
219 - Src/Rcvr Atten & Bias Tees 2-Port
419 - Src/Rcvr Atten & Bias Tees 4-Port
224 - 2nd Src w/Combiner & Switches
423 - Combiner & Switches
??? - Enter Unlisted Option
3. Click on the option that you wish to either enable or remove, and then click
is appropriate.
4. If the desired option is not available in the list, select
will allow you to enter the option number. Enter the option number and follow the instructions on the
display.
8-14
"Accessing the Option Enable Utility" on page
box. A list of available options, similar to the list below, will
Select Desired Option
PNA Series Microwave Network Analyzers
"Contacting Keysight" on
Enable
??? - Enter Unlisted Option
N5227A
8-13.
or
, whichever
Remove
. A dialog box appears that
Service Guide N5227-90001