Repair and Replacement Procedures

Removing and Replacing the Front Panel Assembly

Removing and Replacing the Front Panel Assembly
Tools Required
— T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
— ESD grounding wrist strap
Removal Procedure
Refer to
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
3. With a T-10 TORX driver, remove the 12 screws (item
Before removing the front panel from the instrument, lift and support the
front of the instrument frame.
4. Slide the front panel off of the instrument frame.
5. Disconnect the ribbon cable (item
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests
Keysight N5264A Service Guide
Figure 7-2
for this procedure.
the frame.
board.
that pertain to this removal procedure. Refer to
"Removing the Covers" on page
) from the A1 front panel interface
Table 7-2 on page
7-5.
) from the sides of
7-42.