Keysight N5227A Service Manual page 102

2-port and 4-port pna microwave network analyzer 10 mhz - 67 ghz
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
5. If the observed problem was frequency banded rather than broadband related, set the analyzer frequency
to the center of the problem band. The spectrum analyzer should measure a signal above the network
analyzer setting. For analyzers with DSP version 4.0, the signal is 7.61 MHz above the network analyzer
setting. For analyzers with DSP version 5.0, the signal is 7.44 MHz above the network analyzer setting.
6. If the LO signal is not present but the 50 MHz reference signal from
Reference Outputs"
is present, replace the A15 LO synthesizer board. Refer to
the A4–A17 Boards" on page
7. If the signal is present, reconnect cable W41, and then continue with
Output".
Checking the A25 HMA26.5 Output
1. Refer to the block diagram at the end of this chapter and to
Configuration, Option 400" on page
2. Disconnect W52 (4-port models) or W80 (2-port models) from the A25 HMA 26.5.
3. Connect the spectrum analyzer to the open connector on the A25 HMA 26.5.
4. Refer to
the IMPORTANT notice on page
and observe the spectrum analyzer measurement. For analyzers with DSP version 4.0, an 807.61 MHz
signal should be present. For analyzers with DSP version 5.0, an 807.44 MHz signal should be present.
5. If the observed problem was frequency banded rather than broadband related, set the analyzer frequency
to the center of the problem band. The spectrum analyzer should measure a signal above the network
analyzer setting. For analyzers with DSP version 4.0, the signal is 7.61 MHz above the network analyzer
setting. For analyzers with DSP version 5.0, the signal is 7.44 MHz above the network analyzer setting.
6. If the signal is not present but the signal from
present, replace the A25 HMA 26.5. Refer to
7. If the signal is present, reconnect cable W52 or W80 and then:
• for 2-port models, continue checking with
• for 4-port models, continue checking with
Checking the A26 Splitter Output
1. Refer to the block diagram at the end of this chapter and to
Configuration, Option 400" on page
2. Disconnect W53 and W54, one at a time, from the A26 splitter.
3. Connect the spectrum analyzer to the open connector.
4. Refer to
the IMPORTANT notice on page
and observe the spectrum analyzer measurement. For analyzers with DSP version 4.0, an 807.61 MHz
signal should be present. For analyzers with DSP version 5.0, an 807.44 MHz signal should be present.
5. If the observed problem was frequency banded rather than broadband related, set the analyzer frequency
to the center of the problem band. The spectrum analyzer should measure a signal above the network
analyzer setting. For analyzers with DSP version 4.0, the signal is 7.61 MHz above the network analyzer
setting. For analyzers with DSP version 5.0, the signal is 7.44 MHz above the network analyzer setting.
6. If the signal is not present but the signal from
4-32
7-10.
6-46. Locate the flexible cable W52 at the A25 HMA26.5.
4-23. Set the network analyzer for an 800 MHz CW frequency
"Checking the A15 13.5 GHz LO Synthesizer Output"
"Removing and Replacing the A25 HMA26.5" on page
"Checking the Receiver Group" on page
"Checking the A26 Splitter
6-46. Locate the semi-rigid cables W53 and W54, at the A26 splitter.
4-23. Set the network analyzer for an 800 MHz CW frequency
"Checking the A25 HMA26.5 Output,"
PNA Series Microwave Network Analyzers
"Checking the A14 50 MHz
"Removing and Replacing
"Checking the A25 HMA26.5
"Bottom RF Cables, Standard 4-Port
Output".
"Bottom RF Cables, Standard 4-Port
Service Guide N5227-90001
N5227A
is
7-26.
4-40.
is present, replace

Advertisement