Agilent Technologies N5230A Service Manual page 57

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5230A
Figure 3-12
shows an example of typical verification results with Print Graphs selected in
the Printer Output area of the System Verification dialog box. The printed graphical
results show the following:
• the name of the device measured
• the serial number of the device
• the parameters measured
• Results of the measurements. Labeled as A in
• Data measured at the factory from the verification kit. Labeled as B in
• Upper and lower limit points as defined by the total system uncertainty system.
Labeled as C in
Figure
Figure 3-12 Example of Printed Graphical Verification Results
Service Guide N5230-90014
3-12.
Figure
3-12.
Tests and Adjustments
System Verification
Figure
3-12.
3-27