Agilent Technologies N5230A Service Manual page 97

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5230A
EEPROM Headers
The network analyzer application uses the firmware revision information stored in the pc
board header EEPROM. If the information stored in any EEPROM is incorrect, the
network analyzer may not operate properly.
The following table lists the pc boards in your network analyzer that contain EEPROM
headers. The pc boards are listed by name and part number and the correct firmware
revision code is given for each.
PC Board Name
A5 SPAM
A6 and A8 multiplier
A7 and A9 fractional-N synthesizer
A10 frequency reference
A16 test set motherboard
To view this EEPROM header information on the network analyzer display, select System,
Service, Utilities, View EEPROM Headers in the network analyzer application.
If the information is incorrect for any of the pc boards, contact Agilent Technologies. Refer
to
"Contacting Agilent" on page
Error Messages
SOURCE UNLEVELED: The source ALC circuit on the A16 test set motherboard is running
open-loop. Check the cable connections for W30, W31, and W32 between the A19 SSLAM
and the A16 test set motherboard. Refer to those sections of this chapter that pertain to
troubleshooting the source section of the analyzer.
Service Guide N5230-90014
PC Board Part
Number
E8364-63187
E8364-63182
E8364-63183
E8364-60136
N5230-63078
2-11.
Measurement System Troubleshooting
Memory
Hardware
ID
ID
3
5
3
6
3
7
3
10
3
16
Troubleshooting
Firmware
Revision
A
A
C
B
D
4-27