Calibration Coefficients Test - Agilent Technologies N5230A Service Manual

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5230A

Calibration Coefficients Test

Function of the Test: To verify the uncorrected calibration coefficients of your analyzer.
The calibration coefficients are measured in forward and reverse direction. Refer to
Terms"
in
Chapter 8
measured.
Specification Tested: Uncorrected System Performance
Equipment Used: A calibration kit and a test cable.
Description of the Test:
1. A series of 2-port calibrations are performed. Two full SOLT 2-port calibrations are
performed on each port. Isolation is turned off during each 2-port calibration.
2. A test cable is attached to the first port, and a calibration is performed at the end of the
cable and at the second port. The cable is moved to the second port and another
calibration is performed. Using two calibrations helps to eliminate the unknown
characteristics of the through cable.
3. Each calibration produces 12 error terms.
4. The error terms are used to determine the following characteristics: directivity, source
match, load match, reflection tracking, and transmission tracking.
If the Analyzer Fails this Test:
• If the analyzer fails tracking error terms only, perform
Adjustment" on page 3-38
• Failure of any other error terms indicate a hardware failure. Refer to the appropriate
error term discussion in
to
Chapter 7, "Repair and Replacement Procedures,'
suspected faulty component or assembly.
Service Guide N5230-90014
for error term information relating to the calibration coefficients
and repeat this test.
"Error Terms"
Performance Tests (Agilent N7840A Software Package)
"Receiver Calibration
in
Chapter 8
for a typical cause of failure. Refer
for instructions on replacing the
Tests and Adjustments
"Error
3-33