PNA Series Microwave Network Analyzers
N5230A
Node 718
SOMA 50 Prelevel DAC Node 718 senses the prelevel DAC voltage
applied to the output of the source ALC circuitry upon analyzer startup to
preset the output level of the SOMA 50.
Node Descriptions, A16 Test Set Motherboard, AUX 2
−10 V Reference Node 811 senses the −10 V supply voltage that serves as
Node 811
a measurement reference.
+1.78 V Bias Reference Node 812 senses the +1.78 V bias supply voltage
Node 812
that serves as a measurement reference.
Node 813
Phase Lock IF Detector Node 813 senses the voltage input to the phase
lock IF detector.
+10 V Reference Node 814 senses the +10 V supply voltage that serves as
Node 814
a measurement reference.
−1.25 V Bias Reference Node 815 senses the −1.25 V bias supply voltage
Node 815
that serves as a measurement reference.
+5 V Reference Node 816 senses the +5 V supply voltage that serves as a
Node 816
measurement reference.
Node 817
Logger Drive Voltage Output Node 817 senses the drive voltage applied
to the source ALC log amp.
Node 818
Detector Voltage Output Node 818 senses the output voltage of the
phase lock IF detector.
Service Guide N5230-90014
Troubleshooting
Measurement System Troubleshooting
4-35