Agilent Technologies N5230A Service Manual page 104

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
Node 917
2.250 GHz VCO Compensation Node 417 senses the average voltage
output level of the 2.250 GHz VCO.
Node 918
ALC Level Adjust Node 418 senses the voltage that is used to set the
offset on the heterodyne band ALC.
Node Descriptions, A16 Test Set Motherboard, MA
−15 V Reference Node 516 senses the −15 Vdc supply voltage that serves
Node 516
as a measurement reference.
+9 V Reference Node 517 senses the +9 Vdc supply voltage that serves as
Node 517
a measurement reference.
+15 V Reference Node 518 senses the +15 Vdc supply voltage that serves
Node 518
as a measurement reference.
Node Descriptions, A16 Test Set Motherboard, SSLAM
Node 611
Detector Level Node 611 senses the power level of the internal detector
signal from the A22 switch splitter or the power level from the external
detector from the rear panel.
Node 612
Level Output Node 612 senses the output of the detector log amp in the
ALC.
Node 615
Temperature Compensation Node 615 senses the input voltage applied
to the temperature compensation amplifier for the source ALC.
Node 616
Slope Compensation Node 616 senses the slope compensation.
Node 617
Log Amp Collector Voltage Node 617 senses the collector voltage of the
source ALC log amp.
Node 618
Power DAC Node 618 senses the output voltage of the power DAC. This
output is summed with other compensation signals to control the source
output level. The power DAC has the most influence of the compensation
signals.
Node Descriptions, A16 Test Set Motherboard, AUX 1
Node 711
Log Amp Breakpoint 1 Voltage Node 711 senses the breakpoint voltage
for stage 1 of the source ALC log amp.
Node 712
Log Amp Breakpoint 2 Voltage Node 712 senses the breakpoint voltage
for stage 2 of the source ALC log amp.
Node 714
Log Amp Offset Voltage Node 714 senses the offset voltage for the source
ALC log amp.
+10 V Reference Node 715 senses the +10 V supply voltage that serves as
Node 715
a measurement reference.
Node 716
Log Amp Emitter Voltage Node 716 senses the emitter voltage of the
source ALC log amp.
Node 717
Temperature Compensation Node 717 senses the output voltage of the
temperature compensation amplifier for the source ALC.
4-34
PNA Series Microwave Network Analyzers
Service Guide N5230-90014
N5230A

Advertisement

Table of Contents
loading