Repair and Replacement Procedures

Removing and Replacing the A23 Test Set Motherboard

Removing and Replacing the A23 Test Set Motherboard
Tools Required
• T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
• T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
• 5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
• 5/8 inch nutsetter (torque to 21 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-12
for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
3. Position the analyzer bottom side up.
4. Disconnect ALL ribbon cables (item
motherboard.
5. Remove connector hardware (item
6. Remove 10 screws (item
7. Slide the A23 test set motherboard toward the front of the instrument until the rear panel BNC
connectors are free of the rear panel, then lift the motherboard and remove it from the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to
Table 7-2 on page
7-26
"Removing the Covers" on page
) and ALL wire harnesses (item
) from 11 rear panel BNC connectors.
) from the A23 test set motherboard.
7-56.
PNA Series Microwave Network Analyzers
N5224A and N5225A
7-6.
) from the A23 test set
Service Guide N5225-90001