Removing And Replacing The Lithium Battery - Keysight N5224A Service Manual

2-port and 4-port pna microwave network analyzers
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing and Replacing the Lithium Battery

Removing and Replacing the Lithium Battery
Tools Required
• T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
• T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-26
for this procedure.
1. Disconnect the power cord.
2. Remove the solid state drive (SSD) from the A21 CPU board. Refer to
Solid State Drive (SSD)" on page
3. Remove 22 top cover attachment screws (item
4. Remove the top cover from the A21 CPU board assembly.
5. Remove the battery from the battery holder by lifting it at the open end of the holder and then sliding it
from under the clip (item
6. DO NOT THROW AWAY THE BATTERY. COLLECT IT AS SMALL CHEMICAL WASTE. Refer to
Battery Disposal" on page 1-8
Replacement Procedure
1. Reverse the order of the removal procedure following all instructions included with the new battery.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to
Table 7-2 on page
7-54
7-46.
).
for additional information on battery disposal.
7-56.
PNA Series Microwave Network Analyzers
"Removing and Replacing the A55
).
N5224A and N5225A
"Lithium
Service Guide N5225-90001

Hide quick links: