Error Terms

Error Term Data

The error term descriptions in this section include the following information:
• a table of the error terms
• description and significance of each error term
• measurements affected by each error term
• typical cause of failure for each error term
The same description applies to both the forward (F) and reverse (R) terms.
IMPORTANT Data are listed here as a convenience only. Detailed instrument specifications are listed in
the embedded help in the network analyzer.
If Error Terms Seem Worse than Expected
To verify that the system still conforms to specifications, perform a system verification. Refer to
"Tests and
Adjustments".
Figure 8-5
Error Term Data
Parameter
10 MHz
(All options,
to
all ports)
50 MHz
Directivity
16 dBm
Source
11 dBm
Match
Load Match
11 dBm
10 MHz
to
50 MHz
b
-84 dBm
Crosstalk
a. The data in this table are uncorrected system performance. The values apply over an environmental
temperature range of 25 °C ±5 °C, with less than 1 °C deviation from the calibration temperature.
b. All crosstalk values are typical. Measurement conditions: normalized to a thru, measured with two shorts,
10 Hz IF bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the maximum
power out or the maximum receiver power.
8-10
a
Frequency Range
50 MHz
3.2 GHz
to
to
3.2 GHz
10 GHz
24 dBm
23 dBm
18 dBm
14 dBm
17 dBm
13 dBm
50 MHz
100 MHz
to
to
100 MHz
500 MHz
-90 dBm
-110 dBm
PNA Series Microwave Network Analyzers
N5224A and N5225A
10 GHz
16 GHz
to
to
16 GHz
24 GHz
16 dBm
16 dBm
12 dBm
10 dBm
10 dBm
9 dBm
500 MHz
3.2 GHz
to
to
3.2 GHz
20 GHz
-120 dBm
-122 dBm
Service Guide N5225-90001
Chapter 3,
24 GHz
to
26.5 GHz
16 dBm
8 dBm
8 dBm
20 GHz
to
26.5 GHz
-117 dBm