Source Adjustment - Keysight N5224A Service Manual

2-port and 4-port pna microwave network analyzers
Table of Contents

Advertisement

Tests and Adjustments
Adjustments

Source Adjustment

The source calibration is used to adjust your network analyzer for a flat source power across its full
frequency range. There are differences between each test port; therefore, an adjustment is required for each
port.
Equipment Used for the Source Adjustment
Equipment Type
Power meter
Power sensor, 2.4 mm
Adapter, 2.4 mm (f) to 2.4 mm (f)
Procedure
1. Connect the equipment as shown in
and the power meter.
Figure 3-14
Equipment Setup for the Source Adjustment
2. .Press UTILITY
System
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
3-42
Figure
3-14. Connect a GPIB cable between the network analyzer
Service
Adjustments
, then
, then
PNA Series Microwave Network Analyzers
Model or
Alternate Model or
Part Number
Part Number
E4418B/E4419B
E4418A/E4419A
8487A
None
11900B
85056-60007
, then click Source Adjustment.
Service Guide N5225-90001