Performance Tests; Certificate Of Calibration - Keysight N5224A Service Manual

2-port and 4-port pna microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5224A and N5225A
against the system specifications. If confirmation is successful, the measurement system is capable of
making measurements to the accuracy specified by the graphs of measurement uncertainty.
The procedure consists of calibrating the analyzer with a calibration kit, measuring a set of characterized
devices, and comparing the resultant measured data to the data and uncertainty limits supplied with the
verification kit. The device data provided with the verification kit has a traceable path to NIST. The total
measurement uncertainty limits for the performance verification are the sum of the factory measurement
uncertainties and the uncertainties associated with measuring the same devices on the system being
verified. The difference between the factory-measured data and the verification-measured data must fall
within the total uncertainty limits at all frequencies for the total system uncertainty test to pass.
NOTE
Calibration kits are different from verification kits. Calibration kits are used to determine the
systematic errors of a network analyzer measurement system. Verification kits are used to
confirm system specifications and are not used to generate error correction.

Performance Tests

Performance tests are used to confirm analyzer performance against the "instrument" specifications. If
confirmation is successful, the PNA meets the instrument specifications.
Performance tests are contained in the analyzer's firmware with Options 897 or 898 and are described at
"Performance Tests" on page
An illustrated outline of the performance verification procedure:
• for ANSI/NCSL Z540.3-2006 and ISO/IEC 17025 verification, is shown in
• for non-standards verification, is shown in

Certificate of Calibration

Keysight Technologies will issue a certificate of calibration upon successful completion of system
verification or completion of the performance tests. The certificate of calibration will apply to the "system"
(analyzer, calibration kit, test cables, and any necessary adapters) if the system verification procedure is
used to confirm the system specifications. If the performance tests are used to confirm instrument
specifications, the certificate of calibration will apply to the PNA as an independent instrument. The
equipment and measurement standards used for the tests must be certified and must be traceable to
recognized standards.
NOTE
If you have a measurement application that does not use all of the measurement capabilities
of the analyzer, you may ask your local Keysight Technologies service office to verify only a
subset of the specifications. However, this "limited calibration" creates the possibility of
making inaccurate measurements if you then use the analyzer in an application requiring
additional capabilities.
Service Guide N5225-90001
3-28.
Figure 3-2 on page
About System Verification and Performance Tests
Figure 3-1 on page
3-9.
Tests and Adjustments
3-8.
3-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5225a