Correction Constants; Storing Correction Constants - Keysight N5224A Service Manual

2-port and 4-port pna microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
N5224A and N5225A

Correction Constants

Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM memory. These constants enable
the analyzer to produce accurate, leveled source signals and receive clean test signals.

Storing Correction Constants

After performing any adjustment listed on
page 3-39
in this manual, store the correction constants to a
backup file on the analyzer solid state drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Keysight\Network Analyzer\Service\eebackup.exe
• Run the program.
• Click Backup EEPROM.
• Click Exit when the program has finished.
Service Guide N5225-90001