PNA Series Microwave Network Analyzers
Tests and Adjustments
N5224A and N5225A

Non-Standards Compliant Verification

Non-Standards Compliant Verification
To meet the criteria for non-standards compliant verification, perform the preliminary checks and the
performance tests while stopping to troubleshoot. Refer to
Figure 3-2
for test flow. Troubleshoot and repair
the first problem encountered without continuing to other tests. After you troubleshoot, repair, and adjust,
repeat the last failed portion and generate a new set of data.
Figure 3-2
Non–Standards Compliant Verification Flowchart
Service Guide N5225-90001
3-9