2.6
Assembling the Adapters for the Voltage Input
Terminals
Assembling the 758931 Safety Terminal Adapter
When connecting a measurement cable to a WT1800 voltage input terminal, use the included 758931
Safety Terminal Adapter or the 758923 Safety Terminal Adapter (sold separately). When using the
758931 Safety Terminal Adapter, assemble it according to the following procedure.
Assembling the Safety Terminal Adapter
1.
Remove approximately 10 mm of the covering from the end of the cable and pass the cable
through the internal insulator.
Cable
2.
Insert the tip of the cable into the plug. Fasten the cable in place using the hexagonal wrench.
Hexagonal wrench
Insert the hexagonal wrench into the plug and tighten.
3.
Insert the plug into the internal insulator.
4.
Attach the external cover. Make sure that the cover does not come off.
Note
Once you attach the cover, it is difficult to disassemble the safety terminal adapter. Use care when attaching
the cover.
Below is an illustration of the adapter after it has been assembled.
IM WT1801-03EN
Internal insulator
Attachable cable
Covering: max. diameter 3.9 mm
Core wire: max. diameter 1.8 mm
10mm
Plug
Cable tip
Index
Cover
2-11