2.2 Installing the Instrument
Storage Location
When storing the instrument, avoid the following places.
• Where the relative humidity is greater than 80%
• Where the level of mechanical vibration is high
• In direct sunlight
• Where there are corrosive or explosive gasses
• Where the temperature is 60°C or higher
• Where an excessive amount of soot, dust, salt, or iron is present
• Near a strong source of heat or moisture
• Where water, oil, or chemicals may splash onto the instrument
We recommend that the instrument be stored in an environment where the temperature is between
5°C and 40°C and the relative humidity is between 20% RH and 80% RH.
Installation Position
Desktop
Place the instrument on a flat, level surface as shown in the figure below.
Rubber Stoppers
If the instrument is installed so that it is flat as shown in the above figure, rubber stoppers can be
attached to the feet to prevent the instrument from sliding. Two sets of rubber stoppers (four stoppers)
are included in the package.
Rack Mounting
To mount the instrument on a rack, use a rack mount kit (sold separately).
Name
Rack mount kit
Rack mount kit
A summary of the procedure for mounting the instrument on a rack is given below. For detailed
instructions, see the manual that is included with the rack mount kit.
1.
Remove the handles from both sides of the instrument.
2.
Remove the four feet from the bottom of the instrument.
3.
Remove the two plastic rivets and the four seals covering the rack mount attachment holes on
each side of the instrument near the front.
4.
Place seals over the feet and handle attachment holes.
5.
Attach the rack mount kit to the instrument.
6.
Mount the instrument on a rack.
Note
When mounting the instrument on a rack, allow at least 20 mm of space around the inlet and vent holes to
prevent internal heating.
Make sure to provide adequate support from the bottom of the instrument. The support should not block the
inlet and vent holes.
2-4
Model
Notes
751535-E4
For EIA
751535-J4
For JIS
IM WT1801-03EN