Auxiliary Input (Option); D/A Output And Remote Control (Option) - YOKOGAWA WT1800 User Manual

Precision power analyzer
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

6.8

Auxiliary Input (Option)

Item
Input terminal
Input format
Input resistance
Input connector type
Range
Input range
Line filter
Continuous maximum
allowable input
Maximum common-mode
voltage
Sampling interval
Resolution
Accuracy
Temperature coefficient
6.9

D/A Output and Remote Control (Option)

D/A Output
Item
D/A conversion resolution
Output voltage
Update interval
Number of outputs
Accuracy
Minimum load
Temperature coefficient
Continuous maximum
common-mode voltage
Remote Control
Item
Signal
Input level
IM WT1801-03EN
Specifications
AUX1, AUX2
Analog
Approx. 1 MΩ
Isolated BNC
50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, 10 V, 20 V
±110%
Off, 100 Hz, 1 kHz
±22 V
±42 Vpeak
Approx. 200 kS/s
16 bits
±(0.05% of reading + 0.05% of range)
• Add 20 μV/°C for temperature changes after zero-level compensation or range change.
±0.03% of range/°C
Specifications
16 bits
Each rated value ±5 V FS (maximum of approx. ±7.5 V)
Same as the WT1800 data update interval
(if the waveform display is enabled and the trigger mode is set to Auto or Normal, the data update
interval depends on the trigger operation)
20 channels (the output items can be set for each channel)
±(each measurement function's accuracy + 0.1% of FS); FS = 5 V
100 kΩ
±0.05% of FS/°C
±42 Vpeak or less
Specifications
EXT START, EXT STOP, EXT RESET,INTEG BUSY, EXT HOLD,EXT SINGLE,
EXT PRINT
0 V to 5 V
1
2
3
4
5
6
App
Index
6-15