6.20 General Specifications

Item
Warm-up time
Operating environment
Elevation
Installation location
Storage environment
Rated supply voltage
Permitted supply voltage
range
Rated supply frequency
Permitted supply voltage
frequency range
Maximum power
consumption
External dimensions
(See section 6.21.)
Weight
Battery backup
Safety standards
1
1
Emissions
1
Immunity
1
Applies to products with CE marks. For information on other products, contact your nearest YOKOGAWA dealer.
2
The overvoltage category (installation category) is a value used to define the transient overvoltage condition and includes the rated impulse
withstand voltage. CAT II applies to electrical equipment that is powered through a fixed installation, such as a wall outlet wired to a distribution
board.
3
Measurement Category II (CAT II) applies to electrical equipment that is powered through a fixed installation, such as a wall outlet wired to a
distribution board, and to measurement performed on such wiring.
4
Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas that deteriorates withstand voltage or surface resistivity. Pollution
Degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
5
Use cables of length 3 m or less.
6
Use cables of length 30 m or less.
IM WT1801-03EN
Specifications
Approx. 30 minutes
Temperature: 5°C to 40°C
Humidity: 20%RH to 80%RH
(No condensation)
2000 m or less
Indoors
Temperature: −25°C to 60°C
Humidity: 20%RH to 80%RH
(No condensation)
100 VAC to 240 VAC
90 VAC to 264 VAC
50/60 Hz
48 Hz to 63 Hz
150 VA (when the built-in printer is used)
Approx. 426 (W) × 177 (H) × 459 (D) mm
(When the printer cover is not attached; excluding the handle and other protruding parts.)
Approx. 15 kg (the weight of the main unit with six input elements and all options installed)
Setup parameters and the internal clock are backed up with a lithium battery.
Compliant standard: EN61010-1, EN61010-2-030
The overvoltage category (installation category) is CAT II.
The measurement category is CAT II.
Pollution degree 2
4
Compliant standards
EN61326-1 Class A, EN61000-3-2, EN61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard ( 한국 전자파적합성기준 )
This product is a Class A (for industrial environments) product. Operation of this product in a
residential area may cause radio interference in which case the user is required to correct the
interference.
Cable conditions
• EXT CLK, MEAS. START, motor evaluation function terminals, and AUX input terminals
5
Use BNC cables.
• GP-IB interface connector
Use a shielded GP-IB cable.
• RGB output connector
Use a shielded D-sub 15 pin cable.
• USB port (PC)
5
Use a shielded USB cable.
• USB port (for peripheral devices)
Use a USB keyboard that has a shielded cable.
• Ethernet connector
Use a category 5 or better Ethernet cable (STP).
Compliant standard
EN61326-1 Table 2 (for industrial locations)
Influence in the immunity environment
Measurement input: within ±20% of range
(When the crest factor is set to 6, within ±40% of range.)
D/A output: within ±20% of FS; FS = 5 V
Cable conditions
The same as the cable conditions listed above for emissions.
2
3
5
5
5
6
1
2
3
4
5
6
App
Index
6-21

Advertisement

Table of Contents
loading