6.6 Care

6.6.1 Care Instructions

Do not allow sand or other particles to get into the product. They could degrade the wheel bearings and the
locking mechanism.
Use in saltwater is not permitted.

6.6.2 Cleaning

1) Clean the cushion with warm water and a mild detergent.
2) Remove any spots with a sponge or a soft brush.
3) Rinse with clear water and let the components dry.
Important information on cleaning
Do not use any aggressive cleaners, solvents or hard brushes etc.
Plastic parts, frame parts, and the chassis and wheels can be cleaned with a mild cleaner and a damp cloth.
Dry thoroughly afterwards.
Do not immerse in water. Components are not machine washable.

6.6.3 Disinfection

► Wipe all parts of the wheelchair with a disinfectant.
Important information about disinfecting
Only use water-based disinfectants. Observe the instructions for use provided by the manufacturer.
Thoroughly clean the pads and handles before disinfecting.
7 Maintenance/Repair

7.1 Maintenance

The product bears the CE marking. The manufacturer hereby guarantees that this medical device as a whole con­
forms to the requirements of the European Directive for Medical Devices 93/42/EEC.
The function of the product should be checked before each use.
If defects are found, especially instability of the product or altered driving characteristics as well as problems
with the user's seating position or the stability of the seat, the product should not be used. The specialist deal­
er must be contacted promptly in order to correct these defects.
This also applies if loose, worn, bent or damaged components are identified.
Some maintenance tasks can be carried out to a certain extent by the user at home (see the section "Mainten­
ance Tasks").
The manufacturer also recommends regular maintenance every 12 months by an authorised dealer.
Failure to maintain the product can lead to serious or life-threatening injury to the user of the product.
Service and repairs may only be carried out by authorised dealers or the manufacturer. This will ensure that
only Ottobock spare parts are used for repairs.

7.1.1 Maintenance Intervals

The functions described below must be checked by the user or an attendant at the specified intervals. During these
tasks, the user must not be sitting in the wheelchair: