• Use the cursor keys to adjust the Blinking value. First select KΩ, MΩ, GΩ
or TΩ using the Up Arrow Key ().
• Next, press the Right Arrow Key () to move to the next selection which
is the greater than (>) / less than (<) choice. Use the  to make your
selection.
• Press the  again to move to the Resistance Value Set Point Selection.
You can change this value digit by digit using the  /  combination
working from the highest to lowest digit.
• When finished programming the alarm settings, place the rotary switch to
the desired test function.
• Before initiating a test, press the Yellow 2
to activate the Alarm Notification. The word "ALARM" will appear in the
lower left side of the display.
• Next, press the round yellow Start/Stop button to begin the test. The Alarm
symbol will flash and the buzzer will sound continuously if the Alarm set
point is tripped.
NOTE: If the rotary switch is turned off prior to the next test, the alarm notification
will be deactivated. You will need to press the Yellow 2
button to activate the Alarm Notification when you turn the unit back on before
starting the next test, if alarm notification is desired.
2.5.5

/ SMOOTH Button

First Function -
Selects the desired parameter to be modified - the selected parameter flashes.
It is modified using the
Second Function - SMOOTH
Activates a digital filter for insulation measurement. It only has an effect on the
displayed values (which are smoothed) and not the actual measurements.
For example, this function is useful when the displayed insulation values are highly
unstable, brought about due to a capacitive component in the tested element.
Shop for AEMC products online at:
Megohmmeter Models 1050 /1060
button (see § 2.5.6).
ShopAEMC
www.
button and the Alarm button
nd
button and the Alarm
nd
.com
1.877.766.5412
17