Smooth Button; Button - AEMC 5050 User Manual

Megohmmeter
Hide thumbs Also See for 5050:
Table of Contents

Advertisement

NOTE: If the rotary switch is turned off prior to the next test, the alarm
notification will be deactivated. You will need to press the Yellow 2
button and the Alarm button to activate the Alarm Notification when you
turn the unit back on before starting the next test if an alarm notification
is desired.
2.6.5 ► / Smooth Button
First Function - ►
Selects the desired parameter to be modified. The selected parameter will flash.
The flashing parameter is modified using the
Second Function - SMOOTH
Activates a digital filter for insulation measurement. It only has an effect on the
displayed values (which are smoothed) and not the actual measurements.
For example, this function is useful when the displayed insulation values are
highly unstable, brought about due to a capacitive component in the tested
element.
2.6.6

Button

Changes the flashing parameters displayed, or views R(t) values (see §2.6.3).
As a general rule, two figures (day, month, hour, min., sec., and OBJ:TEST) will
flash.
The ▲ and ▼ functions have a "follow-on" mode (e.g. as soon as the high or low
program limit is reached, the parameter to be modified switches automatically to
the following low or high limit).
First Function ▲
■ Press button briefly to increase the displayed number slowly.
■ Press button for a longer time to increase at a faster rate.
Second Function ▼
■ Press button briefly to decrease the displayed number slowly.
■ Press button for a longer time to decrease at a faster rate.
Unlike all the second functions of other buttons, it is not necessary to press the
button each time to get to the ▼ function. The
and is valid for the ▼ function (only), as long as the user does not deactivate it by
pressing
or another button.
20
button (see §2.6.6).
symbol remains displayed
Megohmmeter Models 5050 & 5060