Advertisement

MEMORY / RS-232

5.1 RS-232 Specifications

The Communication Rate can be set to 300, 600, 1200, 2400, 4800, 9600, or
"Parallel" for printing results on parallel printers via the optional serial-to-parallel
adapter.
This setting is carried out in the SET-UP menu (see § 2.5.7).
Data format: 8 data bits, 1 stop bit, no parity, Xon / Xoff protocol.
Connection to a PC or to a Parallel Printer
Cable, PC RS-232, DB9 F/F 6 ft Null Modem Cable (1060)............. Cat. #2119.45
2
3
5
9-pin female connector
to Model 1060
5.2 Saving / Recalling Values
NOTE: After replacing or completely discharging the battery, the message "Memory
OFF" appears when you try to save a measurement in memory.
To be able to save your measurements, you will need to do the following:
Set the instrument to SET-UP and press the MEM button (the display shows "MEM
All clr"). Press the MEM button for 2s (the display shows "MEM ALL FrEE"). Mea-
surements are now ready to be saved.
First Function - MEM (Save)
Results can be recorded in the instrument's memory and stored along with
addresses defined by an object number (OBJ) and a test number (TEST).
An object represents a 'box' into which we can place 99 tests. Thus, an object can
represent a machine or an installation upon which a certain number of measure-
Shop for AEMC products online at:
ments (insulation, resistance and continuity) are carried out.
34
CHAPTER 5
(MEM/MR Button)
ShopAEMC
www.

(Model 1060)

9-pin female connector
to serial printer
.com
1.877.766.5412
Megohmmeter Models 1050 /1060
2
3
5