Continuity; Resistance (400Kω) - AEMC 1050 User Manual

Megohmmeter
Hide thumbs Also See for 1050:
Table of Contents

Advertisement

Pressing the STOP button stops the measurement. After the measurement has
been stopped the result remains displayed.
It is possible to scroll through all the other results available on the main display
using the R-DAR-PI button. This button can also be used before the measurement
is begun (see § 2.5.3).
If the "time-controlled test" mode has been selected, the R(t) button makes it pos-
sible to access all the saved interval measurements automatically (see § 2.5.2 and
2.5.3).
If the ALARM function is activated, a buzzer is triggered as soon as the measure-
ment crosses the threshold programmed in the SET-UP configuration menu (see
§ 2.5.4).
Display of values after measurement
The following information may be displayed:
R-DAR-PI
Button
Main Display
Resistance
DAR
PI
Capacitance*
R(t)
*Capacitance (µF) measurement is only displayed after the measurement is finished and the
circuit has been discharged.
2.6.3
Continuity (40Ω
• Continuity measurement is carried out on the 40Ω switch position
(with a test current > 200mA up to 20Ω)
• Resistance measurement is carried out on the 400kΩ position
(with a test current < 6mA).
Once the selector switch has been turned to one of these 2 function positions, the
main display indicates - - - - Ω (for continuity) or - - - - kΩ (for resistance) and the
small display indicates the voltage present on the "+" and "–" terminals.
If the voltage is >3V
symbol flashes and the audible alarm beeps (for 2 seconds) signaling that the
measurement was denied. The instrument then goes back to its normal voltage
measurement.
If the voltage is <3V
ment proceeds.
Shop for AEMC products online at:
Megohmmeter Models 1050 /1060
Small display
duration (min. sec)
duration (min. sec)
duration (min. sec)
duration (min. sec)
) / Resistance (400kΩ)
and the START/STOP button is pressed, the
AC/DC
and the START/STOP button is pressed, the measure-
AC
ShopAEMC
www.
V-TIME Button
Small display if the MR button is
pressed (Model 1060)
date, time, test voltage, OBJ : TEST
date, time, test voltage, OBJ : TEST
date, time, test voltage, OBJ : TEST
last voltage
.com
1.877.766.5412
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

1060