3.1 Reference Conditions
Influence Quantity
Temperature
Relative Humidity
Supply Voltage
Frequency Range
Capacitance in parallel with the input resistance
Electric Field
Magnetic Field
3.2 Voltage
Measurement Range: 1 to 1000V
Resolution: 1V
Frequency Range: DC/16 to 420Hz
Accuracy: ± 1% of Reading ± 3cts, AC sinusoidal or DC voltage
Input Resistance: 750kΩ approx
WARNING: This instrument is rated 600V, Cat III, according to EN 61010-1 +
A2 (ed. 95). NEVER use the megohmmeter on electrical conductors rated above
600V.
3.3 Insulation Resistance
Method: Voltage-current method according to EN 61557-2 (ed. 02/97)
Nominal Output Voltage: 50, 100, 250, 500, 1000V
Open-circuit Voltage: ≤ 1.1 x Vn ± 5V (50, 100, 250, 500, 1000V)
Nominal Current: > 1mA
Short-circuit Current: < 6mA
Max. Overvoltage: Vrms max = 1200V
"-" terminals. 660V
Measurement Ranges: 50V: