AEMC 1510 User Manual

Indoor air quality meter
Hide thumbs Also See for 1510:

Advertisement

Quick Links

1510
INDOOR AIR QUALITY
METER
ENGLISH
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for AEMC 1510

  • Page 1 ■ 1510 INDOOR AIR QUALITY METER ENGLISH User Manual...
  • Page 2 Refer to our repair and calibration section at www.aemc.com. Serial #: Catalog #: Model #: 1510 Please fill in the appropriate date as indicated: Date Received: Date Calibration Due: Chauvin Arnoux ® , Inc. d.b.a AEMC ® Instruments www.aemc.com Air Quality Meter Model 1510...
  • Page 3: Table Of Contents

    4.9. Electromagnetic Compatibility ............................18 5. MAINTENANCE ..................................19 5.1. Cleaning ................................... 19 5.2. Battery Replacement ............................... 19 5.3. Firmware Update ................................19 REPAIR AND CALIBRATION ................................20 TECHNICAL AND SALES ASSISTANCE ............................. 20 LIMITED WARRANTY ..................................21 Air Quality Meter Model 1510...
  • Page 4: Precautions

    Thank you for purchasing the Indoor Air Quality Meter Model 1510. For best results from your instrument:  read these operating instructions carefully  comply with the precautions for use WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
  • Page 5: Receiving Your Shipment

    Adapter, Bluetooth USB for Simple Logger II 4-Channel Models, Models 6417, 1510 & L452 ….………………………………………..………………………… Cat. #2126.45 Wall Mount for Models 1510 & L452 (black) ..………..……………………..………..Cat. #2138.61 Calibration Kit for Model 1510 (includes calibration software CD)………....………..Cat. #2138.63 For the accessories and replacement parts, visit our web site: www.aemc.com...
  • Page 6: Getting Started

    1. Temperature and humidity sensors 2. LCD display 3. MODE/temperature units button 4. Backlight/ECO mode button 5. Record button 6. ON/OFF button 7. CO sensor 8. Activate/deactivate buzzer 9. HOLD/Bluetooth enable/disable button 10. MAX/MIN button Air Quality Meter Model 1510...
  • Page 7: Instrument Functions

    Short press enters MAX MIN mode; measurement values continue to be displayed. Second press displays the maximum value. Third press displays minimum value. Fourth press returns to normal measurement operation.  Long press exits MAX MIN mode. Air Quality Meter Model 1510...
  • Page 8: Display

    Steady: connected.   Blinking: Batteries low  Steady: Indicator of external power supply by connection of the mains adapter or of the USB connection. High threshold passed Low threshold passed Air quality/comfort level indicator Air Quality Meter Model 1510...
  • Page 9: Operation

    1D mode. The symbol 1D appears on the display. 2. Press a second time to activate 3D mode. The symbol 3D appears on the display. 3. Press a third time to return to normal operation. Air Quality Meter Model 1510...
  • Page 10: Visible And Audible Warnings

    Arrow in the direction of the overage for each quantity outside the comfort zone. Buzzer sounds (if enabled). Backlight blinks red. Backlight is OFF in ECO mode . For 1D and 3D modes, if the buzzer is active it will sound intermittently when displayed. Air Quality Meter Model 1510...
  • Page 11: Eco (Energy Saving) Mode

    Dry-bulb temperature (°C) 2.4. ECO (Energy Saving) Mode ECO mode saves energy when the Model 1510 measures air quality over a long period. The backlight and the buzzer are deactivated, and CO is measured only once every 10 minutes. This saves the batteries. In addition, the instrument automatically activates “sleep”...
  • Page 12: Min Max Function

    1. Press to activate HOLD. The symbol HOLD appears and the current measurement remains on the screen. 2. Press a second time to return to normal measurement display. Air Quality Meter Model 1510...
  • Page 13: M_Rec (Manual Recording) Function

    . The icon appears on the screen. 2. To deactivate the alarm, press and while holding it down press The alarm buzzer function is disabled in ECO mode. Air Quality Meter Model 1510...
  • Page 14: Error Messages

    To erase memory manually, do the following: 1. Connect the instrument to a computer via the USB cable. 2. Open the computer’s file system. 3. Locate and open the folder named CA1510. 4. Delete any unnecessary files to free memory space. Air Quality Meter Model 1510...
  • Page 15: Dataview

    Control Panel. A green check mark appears next to the name indicating it is currently connected. For instructions about using DataView with the instrument, consult the Data Logger Control Panel Help system. Air Quality Meter Model 1510...
  • Page 16: Specifications

    NOTE: Prolonged exposure to humidity outside the 10% to 80% range may produce measurement bias up to ±3% RH. This bias disappears after 5 days between 68 to 86°F (20 and 30°C) and 40 and 75% RH. Rate of increase of < 0.5% RH/year intrinsic uncertainty: Air Quality Meter Model 1510...
  • Page 17: Influence Of The Temperature On The Relative Humidity Measurement

    ECO mode: approximately 1 year  P_REC 10 minutes mode: 45 days Rechargeable storage batteries can be used, but the period between charges will be shorter. The batteries are not used while the instrument is connected to USB power. Air Quality Meter Model 1510...
  • Page 18: Recording

    Free fall test: Per IEC61010-1 4.8. Compliance with International Standards Compliant with IEC61010-1 safety standard for voltages of 50V in category II. 4.9. Electromagnetic Compatibility Emissions and immunity in an industrial setting compliant with IEC 61326-1. Air Quality Meter Model 1510...
  • Page 19: Maintenance

    Spent batteries must not be treated as ordinary household waste. Take them to an appropriate recycling facility. 5.3. Firmware Update AEMC may periodically update the instrument’s firmware. Updates are available for free download. To check for updates: 1. Connect the instrument to the Data Logger Control Panel.
  • Page 20: Repair And Calibration

    ® ® Contact: Chauvin Arnoux , Inc. d.b.a. AEMC Instruments Phone: (800) 945-2362 (Ext. 351) • (603) 749-6434 (Ext. 351) Fax: (603) 742-2346 E-mail: techsupport@aemc.com Air Quality Meter Model 1510...
  • Page 21: Limited Warranty

    LIMITED WARRANTY Your AEMC instrument is warranted to the owner for a period of two years from the date of original ® purchase against defects in manufacture. This limited warranty is given by AEMC Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused, ®...
  • Page 22 NOTES: Air Quality Meter Model 1510...
  • Page 23 NOTES: Air Quality Meter Model 1510...
  • Page 24 02/18 99-MAN 100460 v1 ® ® Chauvin Arnoux , Inc. d.b.a. AEMC Instruments 15 Faraday Drive • Dover, NH 03820 USA • Phone: (603) 749-6434 • Fax: (603) 742-2346 www.aemc.com...