Executing Coordinated Measurements; Figure 7-4. State Trace Specification With Trigger In Row "B - HP 1630A Operating And Programming Manual

Logic analyzer
Hide thumbs Also See for 1630A:
Table of Contents

Advertisement

Interactive State and Timing Measurements
Model 1630A/D/G
Executing Coordinated Measurements
7-4
1 . Press the RUN key. The analyzer will find the state you specified as its trigger, and complete a state
trace . If the analyzer does not find its state trigger (TIME+0011 on the 16 address lines of the address
bus), the message "Waiting for State Trigger" will be on screen . In this case, press the STOP key . The
message "Trace Aborted" will be on screen, along with a listing of the last eight transactions before you
pressed the STOP key .
2. Write down one of the executions shown on the display under the ADDRS label (such as TIME+001 A) .
3. Press the TRACE key .
4. Move the cursor down into the row "b" line .
5. Enter the module name and offset value from step 2 (use keyboard NEXT[] and hexadecimal keys) . See
figure 7-4 .
1E' 1 .
1
Trace Specification______________________________-_____
1jjjg1
Trace Mode
Master
In sequence
Trace at ®Occurrences of
while storing
then store
Label>
ADDRS
ADDR1
DRTR
CMTRL
WR-RD
MEMIO
Base >
[TIME 155 ] 1
N93fl
XX
XXXX
(XXXXXI
[KEYBO) 555EI
51[18
XX
XXXX
[XXXXXI
[ abs I XXXX
XXXX
XX
XXXX
[XXXXXI
abs IEXXXXEXXXXMXXmXXXXIIXXXXXI
Figure 7-4 . State Trace Specification With Trigger In Row "b"
6. Move the cursor up into the trigger field and press the NEXT[] key to change[ a ]to[ b ] . This sets the
analyzer to trigger when it finds the state in row "b" (the state from step 2) .
7. Press the RUN key. The analyzer will find the state specified as its trigger, and complete a state trace. It
will show the message "Trace Complete" at the top of the display . The state selected as the trigger will
be on line 0000.
8. Press the LIST key several times. Notice that pressing any of the INPUT DISPLAYS and OUTPUT
DISPLAYS keys switches between the state analyzer and timing analyzer .
9. Press the FORMAT key. Press it a second time, if necessary, to place the [Timing] format specification
on screen .
10 . Check the display of the Activity line . If bit 2 in the timing pod shows no activity, connect bit 2 (the third
line) from the timing probe to any active node in the target system .
11 . Press the TRACE key. See figure 7-5 .
12. Move the cursor into the Pattern field and enter a pattern of XXXX1X1X (with a binary display base) .

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1630g1630d

Table of Contents