Installation
Model 1630A/D/G
LINE VOLTAGE SELECTION
2-2
BEFORE CONNECTING THIS INSTRUMENT to line (Mains) voltage, be sure the line
voltage switch is set correctly and that the proper fuse is installed .
If the line fuse burns out, do not replace it until the cause for failure has been determined
and repaired by a qualified service person . Replacing this fuse in a damaged instrument
can cause additional damage .
AC POWER CABLE
OPERATOR'S MAINTENANCE
WARNING
BEFORE CONNECTING THIS INSTRUMENT, the protective earth terminal of the
instrument must be connected to the protective conductor of the (Mains) power cord.
The Mains plug must be inserted in a socket outlet provided with a protective earth
contact. The protective action must not be negated by the use of an extension cord
(power cable) without a protective conductor (grounding) . Grounding one conductor of
a two conductor outlet does not provide an instrument ground .
The only instrument maintenance an operator needs to perform is to occasionally clean the display screen and
the front panel keyboard . When cleaning the 1630A/D/G, caution must be used with which cleaning agents
are used . Use a mild soap and water solution . If a harsh soap or solvent is used, the water-base paint finish will
be damaged .
SITE SELECTION
REAR PANEL ADDRESS SWITCH
CAUTION
Use care when cleaning the front panel keyboard . Water can damage the keyboard
circuitry if it seeps under the keys.
Do not use chemical cleaning agents or abrasive cleaners that may damage the plastics in
this instrument . Recommended cleaning agents are isopropyl alcohol, kelite (1 part
kelite, 20 parts water), or a solution of 1% mild detergent and 99% water .
Place the logic analyzer on a clean workspace which has adequate ventilation .
The dip (dual-inline-package) switch on the rear panel of the logic analyzer is used to configure the analyzer
for operation on the HP-IB and/or HP-IL interface buses, and to execute the self-test routine . The first five
switches set the address of the analyzer for HP-113 operation. Switch six executes the analyzer self-test when
set to "1" . The positions of switches 7 and S determi ne the analyzer configuration on the HP-I B/HP-I L interface
buses.