Figure 5-3. Example State Time Interval Overview Trace Specification - HP 1630A Operating And Programming Manual

Logic analyzer
Hide thumbs Also See for 1630A:
Table of Contents

Advertisement

Model 1630A/D/G
Performance Analysis Measurements
'n sequence St,art timer on
op . timer on
iahel> ADDR
DATA
STHT1
STET
B .?se " ® ®
--
----------- ------------- Ranges --------------------------
- Range 1
= Range 2 <
c --Range 3 <
<= Range 4
= Range 5 <
= Range 5 <:
<= Range 7
= Pange 8
Figure 5-3. Example State Time Interval Overview Trace Specification
NOTE
Use ROLL up key to see entire display.
Sequence Terms :
one or two sequence terms may be defined . Refer to State Trace discussion for more
information .
Occurrence Counter:
refer to State Trace discussion .
Start Term :
the matching pattern to be applied to start the counter. This value can be as simple
as an [any state] or as complex as four OR'd states .
Stop Term :
the matching pattern to be applied to stop the counter. This value can be as simple
as an [any state] or as complex as four OR'd states .
Interval Ranges :
the ranges in which the collected data should be grouped into . The left field
determines the minimum value of the range and the right field determines the
maximum value of the range . These values can be as small as 0 As or as large as
9999s.
Readouts Available :
Time Interval Histogram only .