Figure 8-3. 1630G System [Peripherals] Specification Menu With Internally Stored Information - HP 1630A Operating And Programming Manual

Logic analyzer
Hide thumbs Also See for 1630A:
Table of Contents

Advertisement

The Peripherals Menu
Model 1630A/D/G
Sy-em Specification ------------------------------------------
CI ool: =:et, :
Print non-waveform men
Rear panel port :
Beeper :
HP-IL : Eternal Controller
HF'-IB : 1S30G is controller
Logic . analyzer HP-IB address : 00
Printer HP-IB address :
Disc HP
-
1B address :
02 Ho
Disc
Autoload tn "." erse assembler from ir~t.ernal storag
Hi.rtoloa~d configuration from internal storage :
S-e configuration in internal _storage :
i .-J
Figure 8-3. 1630G System [Peripherals] Specification Menu
With Internally Stored Information
Autoload Inverse Assembler From Internal Storage:
using the PREV[], NEXT[ ]keys can be toggled between [yes] and [no] . Anytime the
selection to [yes] is made the inverse assembler that is stored in memory is loaded to
be active . If [yes] is selected the inverse assembler is automatically loaded the next
time the instrument is powered up .
Autoload Configuration From Internal Storage:
using the PREV[], NEXT[] keys can betoggled between [yes]and [no] . Anytimethe
selection to [yes] is made the previously stored configuration in memory is loaded to
be active . If [yes] is selected the configuration is automatically loaded the nexttime
the instrument is powered up .
8-4
NOTE
If "Autoload inverse assembler from internal storage" and
"Autoload configuration from internal storage" are both set
to [yes] at power up, the inverse is loaded and then the
configuration is loaded . If the configuration does not
contain the same format specification as the inverse
assembler, the inverse assembler may not function properly .
Save Configuration In Internal Storage :
using the PREV[], NEXT[] keys [yes] can be selected, at which time the current
configuration is stored in internal storage, then the selection returns to [no] .
If the internal storage is not operating properly when power is applied or when it is attempted to save a
configuration in internal storage, the message "Internal storage failed power-up" will replace the peripherals
menu entries "Autoload inverse assembler from internal storage" and °Autoload configuration from internal
storage" . The peripherals menu entry "Save configuration in internal storage" will still be displayed. If [yes] is
selected, a reset will be attempted on the internal storage before configuration storage commences. During
the reset, the message "WAIT resetting internal storage" will be displayed . If the reset fails, the message
"Internal storage failed power-up" will remain on-screen . If the reset is sucessful, the current configuration will
saved in internal storage. The "Internal storage failed power-up" message will be replaced with "Autoload
configuration from internal storage" .