AP Elevator Tilt Motor Cabling

For this task the following tools are necessary:
• 1 Allen key 3 mm.
This section describes how the Tilt Motor Cabling is
mounted.
1.
The first cable bracket must be mounted with the ca-
ble jacket protruding approximately 5 mm / 0.2". See
fig. 131.
2.
The cable two brackets on the AP-elevator are
mounted with one screw each. See fig. 131 - 132.
Use a Torque wrench to tighten the screws. Tighten-
ing torque 1.2 Nm / 0.9 Ft/lbs.
3.
Based on the seat depth, the cable either requires
five or four cable brackets on the right hand side of
the seat. If the seat depth is set between 370 – 520
mm (15" - 21"), the cable is mounted in five cable
brackets. See fig. 133.
Figure 131. The cable bracket is mounted with the cable
jacket protruding approximately 5 mm / 0.2".
Figure 132. Cable bracket mounted on the AP Elevator
Tilt Motor.
Figure 133. Tilt Motor Cable mounted with five cable
brackets (Seat depth 370–520 mm (15"-21")).