Permobil F5 Corpus Service Manual page 170

Hide thumbs Also See for F5 Corpus:
Table of Contents

Advertisement

Permobil F5 Corpus
L
WARNING!
Diagnostics should only be performed by persons with sound knowledge of the wheelchair's electronic control
system. Incorrect or poorly performed repair works may make it dangerous to use the wheelchair. Permobil accepts
no liability for any personal injury or damage to the wheelchair and its surroundings that occurs on account of
incorrect or poorly performed repair work.
Example of error messages and remedies
170
EXAMPLE OF ERROR MESSAGES AND REMEDIES
1 Lamp — Low battery voltage
Check the condition of the batteris. Check the contact between the bat-
tery and the control unit.
2 Lamps — Failure in left drive motor
Check the connection of the left drive motor.
3 Lamps — Short-circuit in left drive motor
Check the left drive motor's contacts and cables.
4 Lamps — Failure in right drive motor
Check the connection of the right drive motor.
5 Lamps — Short-circuit inright drive motor
Check the right drive motor´s contacts and cables.
6 Lamps — Battery charger connected
Disconnect the battery charger.
7 Lamps — Joystick error
Check that the joystick has not been moved when the wheelchair is
started.
8 Lamps — Control system error
Check the contacts to the output stage.
9 Lamps — Failure in brake circuit
Check the contacts to the magnetic brakes.
10 Lamps — High battery voltage
Check the battery and the contacts between the battery and the output
stage.
7+5 Lamps — Communication error
A communication error has been indicated. Check that the cable to the
control panel is not damaged and is correctly inserted.
Actuator indicator — Actuator error
An actuator error has been indicated. If the wheelchair is fitted with
more then one actuator, check which one is not working. Check the ac-
tuator´s cable connections.