Remote Control

5.3.16 "MSR" Command

The 'MSR' command it is the only read/write command for LiAM and allows
reading and setting of the value of the slew rate, in A/s, of the power supply.
The reading command is structured as follows:
and the related reply format is:
where:
- value is actual slew-rate value for the power supply [A/s], that is the content of
the EEPROM cell 30, with a 5-digit precision.
The value of the slew-rate can also be set, between lower limit of 0 A/s and
upper limit of 100 A/s, using the MSR command and adopting the following syntax:
where:
- set_value is desired slew-rate value [A/s].
The MSR setting command generates, upon reception, an acknowledgment –
i.e. '#AK\r' - reply if the value is correctly set and with a '#NAK\r' if:
 the set value is incorrect (e.g. out of limits);
 the module is still performing a ramp (it is necessary to wait for the power
supply to end the previous ramp).
Examples:
MSR reading example:
MSR:?\r
MSR setting example with a value out of limits:
MSR:13005\r
MSR setting example to a correct value:
MSR:5.5\r
MSR:?\r
#MSR:value\r
MSR:set_value\r
LiAM User's Manual
#MSR:15.00000\r
#NAK\r
#AK\r
70