5.3.9 "MRP" Command

The 'MRP\r' command returns the value of the positive power rail voltage, i.e.
positive DC Link. Readback values have a unipolar 12-bit resolution.
Replies from the LiAM power supply unit to this command are in the following
format:
where:
 value is the measured DC Link voltage [V].
Even if the internal ADC dedicated to this purpose has a 12-bit resolution, this
value is presented to the user only with a 100 mV resolution.
The MRP command, being a reading command, returns a response in any
module condition.
Examples:
MRP example when the rail voltages are disabled:
MRP\r
MRP example when the positive rail is at mid-level:
MRP\r
63
#MRP:value\r
Remote Control
#MRP:0.0\r
#MRP:55.3\r