5.3.13 "MRV" Command

The 'MRV\r' command returns the readback value of the power supply actual
output voltage, measured at the LiAM module output terminals.
As for the output current, voltage readback values have an 18-bit resolution
(17-bit + sign) and they are presented with a 5-digit precision.
Replies from the power supply LiAM controller to this command are in the
following form:
where:
 value is the output voltage readback [V], measured at the module output
terminals.
The MRV command, being a reading command, returns a response in any
module condition.
Examples:
MRV example when the module is OFF:
MRV\r
MRV example when the module is ON and regulating output current:
MRV\r
67
#MRV:value\r
Remote Control
#MRV:+0.00012\r
#MRV:-28.34563\r