Setting the Clamp Level
A preset clamp value (abnormal value) can be output in cases where the light density is
temporarily excessive or insufficient (i.e. measurement is not possible) due to defects or
holes, for example, in the measurement target.
FUN mode-[I/O]-[ANALOG]-[CLAMP]
Setting value
At current output: MIN (approx.2 mA), MAX (approx.25 mA, default value), 4 to 20 mA (in 1
At voltage output: MIN (approx.-11 V), MAX (approx.11 V, default value), -10 to 10 V (in 1 V
94
I/O Settings
I/O/OUT MIN
[ 0]
OK
mA increments)
increments)
3
Input
the
corresponding to output MIN, and
press the SET key.
Note
To change numerical values, use the
↑UP
key/↓DOWN
change the number of digits use the
←LEFT key/→RIGHT key.
4
Repeat steps 1 to 3 to set output
MAX.
correction
value
key,
and
to
ZG User's Manual