Omron ZG User Manual page 87

2d profile measuring sensors
Hide thumbs Also See for ZG:
Table of Contents

Advertisement

Setting the correction method
Sets correction ON/OFF in both the height and position directions.
FUN mode-[MEAS]-[CORECT]-[HGT POSN]-[METHOD]
Setting value
NONE (no correction)
HEIGHT (height
correction)
POSITION (position
correction)
HEIGHT&POS (height/
position correction)
ZG User's Manual
MEAS/REF.POINT
POS. EDGELV
[150]LV
SET
MEAS/REF.POINT
SEARCH
DIRECTION
SELECT
1L→R 2R→L
MEAS/REF.POINT
EDGE SELECT
SET
Description
The position is not corrected. (default value)
Correction is performed in the height direction.
Correction is performed in the position direction.
Correction is performed in both the height and position directions.
4
To correct shift in the position
direction, align the line of the
edge level to be detected.
Up/Down key: Moves the cursor.
SET: Applies the setting.
ESC key: Cancels the setting.
5
Select the direction in which
edges are searched.
Select [L→R] (forward direction) or
[R→L] (reverse direction).
6
A cross cursor is displayed at the
reference position. Press the ←/→
key to select the edge, and press
the SET key.
Measurement Settings
85

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zg seriesZg-wds70Zg-wds22Zg-wds8tZg-wds3t

Table of Contents