Setting Scaling (Correction Processing)

If scaling is set, differences between measurement values and actual sizes, that occur
due to the color, material or other factors of the measurement target, can be corrected.
There are two scaling setup modes, "automatic setting" and "manual setting." In the
automatic setting mode, actual measurement is performed, and in the manual setting
mode, the correction values are set manually.
(default value: OFF)
Measurement value
5 mm
4.95 mm
Important
The settings below return to the default settings when scaling is set. Set these items after
scaling settings have been completed.
• Zero reset
Measurement items
2-pt step, 3-pt step, edge
width
Height, edge position
Angle, cross-sectional
area
ZG User's Manual
This error is corrected.
Actual size value
5 mm
Setup Method
Automatic setting: one-point scaling
Measurement is performed once, and the correction
value for the measurement value is set.
Automatic setting: two-point scaling
Measurement is performed at two positions, and
correction values are set for those measurement
values.
Manual setting
Input and set coefficients and offset values directly as
numerical values.
Reference
p.72
p.72
p.73