Page 1
INSTRUCTION MANUAL MODEL 200A NITROGEN OXIDES ANALYZER TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (T-API) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.com WEB SITE: www.teledyne-api.com 02246 REV. F1 Copyright 2005 API Inc. 19 August 2005...
Page 2
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 TABLE OF CONTENTS SAFETY MESSAGES ......................II TABLE OF CONTENTS....................III LIST OF FIGURES .......................VII LIST OF TABLES ....................... VIII 1 HOW TO USE THIS MANUAL ................1-1 2 GETTING STARTED ....................2-1 2.1 U...
Page 4
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.3.1 Configuration Information (CFG)..................5-14 5.3.2 Automatic Calibration (AutoCal) ..................5-14 5.3.3 Data Acquisition System (DAS)................... 5-14 5.3.4 Range Menu .......................... 5-16 5.3.5 Password Enable ........................5-20 5.3.6 Time of Day Clock........................ 5-20 5.3.7 Diagnostic Mode........................
Page 5
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.9 Summary of Quality Assurance Checks ................7-29 7.8.10 ZERO and SPAN Checks ....................7-30 7.8.11 Recommended Standards for Establishing Traceability............. 7-31 7.8.12 Certification Procedures of Working Standards ..............7-32 7.9 C...
Page 6
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.2 RS-232 Communications ...................... 9-38 9.3.3 Voltage/Frequency (V/F) Board ................... 9-41 9.3.4 Status/Temp Board........................ 9-47 9.3.5 Power Supply Module ......................9-49 9.3.6 Ozone Generator ........................9-54 9.3.7 Flow/Pressure Sensor......................9-57 9.3.8 NO...
Page 7
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 LIST OF FIGURES 2-1: R & C ........2-3 IGURE EMOVAL OF HIPPING CREWS HECK FOR ORRECT OWER 2-2: R ........................2-4 IGURE ANEL 2-3: I .................... 2-5 IGURE...
Page 8
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 LIST OF TABLES 2-1: F ................2-12 ABLE INAL EST AND ALIBRATION ALUES 2-1: F )............2-13 ABLE INAL EST AND ALIBRATION ALUES ONTINUED 4-1: F LED' ....................4-5...
Page 9
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7-15: A ..........7-26 ABLE UTOMATIC ALCULATION OF ONVERTER FFICIENCY 7-16: D ........7-28 ABLE EFINITION OF EVEL EVEL ERO AND HECKS 7-17: A ................7-29 ABLE CTIVITY ATRIX FOR...
Page 10
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK...
Analyzer Instruction Manual, 02246, Rev. F1 1 HOW TO USE THIS MANUAL The Model 200A has been designed to provide serviceability, reliability and ease of operation. The M200A's microprocessor continually checks operating parameters such as temperatue, flow, and critical voltages. The instruments modular design uses captive screws to facilitate repair and ease of access.
Page 12
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK...
Teledyne API. CAUTION To avoid personal injury, always use two persons to lift and carry the Model 200A. Before operation, it is necessary to remove the shipping hold-down screws. Remove the instrument cover, then refer to Figure 2-1 for screw location.
Page 14
Vent pump exhaust to well ventilated area at atmosphere pressure FIRE or EXPLOSION HAZARD. Do not use charcoal treated with halogen compounds – use only Teledyne API P/N 00596 charcoal. Wait at least 5 minutes after turning off pump before removing ozone scrubber WARNING.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 2.3 Initial Operation Turn on the instrument power. The display should immediately light, displaying the instrument type (M200A) and the computer's memory configuration. If you are unfamiliar with the M200A, we recommend that you read the overview Section 4 before proceeding.
Page 19
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 When the instrument is warmed up, re-check the TEST functions against Table 2-1. All of the readings should compare closely with those in the Table. If they do not, see Section 9.1.1.
Page 20
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure Step Number Action Comment Input Zero gas Allow Zero gas to enter the sample port on the rear of the instrument.
Page 21
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Step 3 - Review the quality of the calibration: Calibration Quality Check Procedure Step Number Action Comment Scroll the TEST The SLOPE value for NOx should be 1.0 ± 0.3. If the value is function menu until not in this range, check Section 7.10 or 9.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 2-1: Final Test and Calibration Values Observed Test Values Units Nominal Range Reference Section Value RANGE 50-20,000 5.3.4 STABILITY < 0.2 9.1.1, Table 9-1, (Zero Gas) 9.2.5 SAMP FLW cc/min 500 ±...
Page 25
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 2-1: Final Test and Calibration Values (Continued) Observed Parameter Units Nominal Range Reference Section Value NO Span Conc 20 - 20,000 Table 7-3 Span Conc 20 - 20,000...
Page 26
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK 2-14...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 3 SPECIFICATIONS, AGENCY APPROVALS, WARRANTY 3.1 Specifications Ranges In 1ppb increments from 50ppb to 20,000ppb Range Modes Single, Independent, AutoRange Noise at Zero 0.2 ppb RMS Noise at Span <0.5% of reading above 50 ppb...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 3.2 EPA Equivalency Designation Advanced Pollution Instrumentation, Inc., Model 200A Nitrogen Oxides Analyzer is designated as Reference Method Number RFNA-1194-099 as defined in 40 CFR Part 53, when operated under the following conditions: Range: Any range from 50 parts per billion (ppb) to 1 ppm.
All replacement parts and repairs are warranted for 90 days after the purchase. If a Product fails to conform to its specifications within the warranty period, Teledyne API shall correct such defect by, in Teledyne API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product.
Page 30
Analyzer Instruction Manual, 02246, Rev. F1 TERMS AND CONDITIONS All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.
Analyzer Instruction Manual, 02246, Rev. F1 4 THE M200A NO ANALYZER 4.1 Principle of Operation The Teledyne API Model 200A Analyzer is designed to measure the concentration of nitric oxide [NO], total oxides of nitrogen [NO ] and, by calculation, nitrogen dioxide [NO ]. The...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 4.2 Operation Summary 4.2.1 Sensor Module, Reaction Cell, Detector The sensor module (Figure 9-9) is where the chemiluminescent reaction takes place and where the light from the reaction is detected. It is the most complicated and critical sub-assembly in the entire analyzer.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 4.2.3 Computer Hardware and Software CPU Board The M200A Analyzer is operated by an NEC V40 microprocessor. Computer communication is done via 2 major hardware assemblies, the V/F board and the front panel display/keyboard. The computer's multitasking operating system allows it to control the instrument, monitor test points, generate analog outputs and provide a user interface via the display, keyboard and RS-232 port.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 The center field displays TEST values. The TEST functions allow you to quickly access many important internal operating parameters of the M200A. This provides a quick check on the internal health of the instrument.
Connect the exhaust (Figure 2-3) to a pump with <3 m of 1/4" PTFE tubing. Failure to meet the performance specifications will result in poor analyzer performance, damage to the pump, damage to the analyzer, and may jeopardize warranty repairs. Teledyne API strongly recommends that the factory supplied pump be used with the M200A.
Page 37
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 NOTE On vacuum vs absolute pressure: Many vacuum gauges read relative to ambient pressure, therefore a reading of 25" of mercury (Hg) at sea level (which would give an absolute pressure of about 5"...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 4.2.8 Ozone Generator Because of the instability of ozone, it is necessary to generate this gas inside the analyzer. The ozone generation module consists of a high voltage AC supply and silent discharge tube and permeation type air drier.
Page 40
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK 4-10...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5 SOFTWARE FEATURES The M200A control software has two major operating modes. The SAMPLE mode is the normal mode when the instrument is taking data. The software menu that covers the SAMPLE mode is diagrammed in Figure 5-1.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.1.1 Sample Menu Table 5-1: M200A Sample Menu Structure Menu Level Reference Level 1 Level 2 Level 3 Level 4 Description Section TEST Test functions Table 9-1 5.2.1, TST>...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.1.2 Set-Up Menu Table 5-2: M200A Setup Menu Structure Setup Menu #1 Reference Level 1 Level 2 Level 3 Level 4 Description Section CFG is primarily used for 5.3.1...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-3: M200A Menu Structure - Setup Menu #2 Setup Menu #2 Reference Level 1 Level 2 Level 3 Level 4 Description Section RANGE Range control menu 5.3.4 MODE Range mode select –...
Page 47
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Setup Menu #3 Reference Level 1 Level 2 Level 3 Level 4 Description Section VARS Internal variables 5.3.9, 9.2 PREV, PREV, NEXT scroll up and 5.3.9, 9.2 NEXT, down through the VARS menu.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.2 Sample Mode The M200A software has two major operating modes, the SAMPLE mode and the SETUP mode. The SAMPLE mode is the normal mode when the instrument is taking data. The SETUP mode is used when the instrument is initially installed, or when problems indicate a need for diagnostics.
Page 49
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Stability The instrument noise is computed using the standard deviation of the last 10 minutes of data, with the value being computed at the end of each NO/ NO cycle.
Page 50
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 High Voltage Power Supply (HVPS) The HVPS reading is a measure of the scaled-up HVPS programming voltage. The voltage used to set the HVPS output is generated on the Preamp board. Its value is between 0 and 1 volt, corresponding to a voltage of 0 to 1000 volts out of the HVPS.
Page 51
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Moly Temperature The moly temperature is controlled by the computer. The nominal set-point is 315 ± 5° C. The temperature sensor inside the moly is a type-J thermocouple. The thermocouple amplifier is located on the STATUS/TEMP board.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.2.2 CAL, CALS, CALZ, Calibration Functions The calibration and zero/span checking of the M200A analyzer is treated in detail in Section 7, Table 7-1 summarizes types of calibration. The basic function of each of these keys is described here.
Page 53
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.2.2.5 Formula Values The slope and offset terms should be checked after each calibration. The values for these terms contain valuable information about the internal health of the analyzer. The range of acceptable values and their meanings is given in Section 7.10.
5.3.1 Configuration Information (CFG) This menu item will tell if the installed software has factory special features or other non- standard features. If you call Teledyne API service you may be asked for information from this menu. 5.3.2 Automatic Calibration (AutoCal) The AutoCal feature allows the M200A to automatically operate the Zero/Span Valve or IZS option to periodically check its calibration.
Page 55
The attributes of the data collecting channels can be changed, and new data collectors can be defined. A comprehensive guide to programming the DAS interface is available separately from Teledyne API, order “DAS Manual” p/n 02837. Table 5-5 is an example of changing a DAS channel attribute.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-5: DAS Data Channel Editing Step Action Comment Press SETUP-DAS-EDIT Enter DAS menu to edit Data Channels Select Data Channel to edit Press PREV/NEXT Press EDIT Enter the Edit menu for the selected Data Channel Scroll through Data Channel properties until RS-232 Press SET>...
Page 57
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.3.4.1 Single Range This range option selects a single range for all output channels (NO, NO , NO ) of the M200A. To select Single Range press SETUP-RNGE-MODE-SING, then press ENTR. To set the value for the range press SETUP-RNGE-SET, enter the full scale range desired from 50 ppb to 20,000 ppm, then press ENTR.
Page 58
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 To use the Dilution feature: SELECT UNITS For stack measurement, select PPM units. To set units, press SETUP-RANGE-UNIT-PPM. Press ENTR after the unit selection is made, then EXIT to return to upper level menus.
Page 59
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.3.4.5 Concentration Units The M200A can display concentrations in ppb, ppm, ug/m3, mg/m3 units. Concentrations displayed in mg/m and ug/m use 0° C, 760 mmHg for STP. Consult your local regulations for the STP used by your agency.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.3.5 Password Enable If password protection is enabled, a password is required to access calibration or setup menus. In the VARS menu a password is always required. To enable passwords press SETUP-PASS-ON.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.3.9 Variables Menu (VARS) This menu enables you to change the settings on certain internal variables. The VARS Table 9-5 is located in the Troubleshooting Section 9.1.4. 5.3.10 M200A Operating Modes The M200A has 2 main operating modes which were discussed in earlier in this section, namely SAMPLE and SETUP modes.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-7: M200A Operating Modes Mode Description ZERO CAL D Automatic dynamic zero calibration ZERO CAL A Automatic zero calibration ZERO CAL R Remote zero calibration ZERO CAL M...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-8: Status Output Pin Assignments Output # Pin # Definition Condition ZERO CAL CLOSED IN ZERO CAL SPAN CAL CLOSED IN SPAN CAL FLOW ALARM CLOSED IF FLOW WARNING...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.5.1 Setting Up the RS-232 Interface The baud rate is set from the front panel by SETUP-MORE-COMM-BAUD. Select the baud rate appropriate for your application, 300, 1200, 2400, 4800, 9600, or 19,200. It is important to note that the other device must have identical settings in order for the communications to work correctly.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-10: RS-232 Switching From Terminal Mode to Computer Mode Function Control-T (ASCII 20 decimal) Switch to terminal mode (echo, edit) Control-C (ASCII 3 decimal) Switch to computer mode (no echo, no edit) If the command line doesn't seem to respond to keystrokes or commands, one of the first things you should do is send a Control-T to switch the command line interface into terminal mode.
Page 67
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Security Feature The RS-232 port is often connected to a public telephone line which could compromise instrument security. If the LOGON feature is implemented the port has the following attributes: A password is required before the port will operate.
Communication Protocols The RS-232 port can communicate in the following protocols. 1. Teledyne API protocol – some of the features of this protocol are covered in Section 5.5.2. A comprehensive manual is available separately as p/n 01350 RS-232 Programming Notes.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-12: RS-232 Command Summary Commands Definition ? [id] Print help screen. ID is an optional instrument ID number T [id] LIST Print all active test messages T [id] LIST name or Print single test message "name"...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-14: RS-232 Interface Command Types First Character Message Type Calibration Diagnostic Test measurement Variable Warning There are 5 different types of messages output by the M200A. The meanings of the various messages are discussed elsewhere in the manual.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 5.5.3 TEST Commands and Messages Table 5-15: RS-232 Test Messages Name Message Description RANGE RANGE=xxxxx PPB Analyzer range NOXRANGE NOX RNG=xxxxx PPB Indep. Range for NO channel NORANGE NO RNG=xxxxx PPB Indep.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 The T command lists TEST messages. Examples of the T command are: T LIST Lists all active test messages T LIST ALL Lists all test messages T CONVTEMP Prints the temperature of the moly converter...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Whenever a warning message is reported on the analyzer display, if the RS-232 interface is in the normal mode(i.e. not in quiet mode) the warning message is also sent to the RS-232 interface.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 5-18: RS-232 Calibration Examples Action RS-232 Commands Comments Zero Calibration C ZERO Z/S valves switched to admit C COMPUTE ZERO zero gas. Zero cal in Single C EXIT Range mode.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 The following is a summary of the Diagnostic commands. Table 5-20: RS-232 Diagnostic Command Summary Command Description D [id] LIST Prints all I/O signal values. See Table 9-4 for Sig I/O definitions.
Page 78
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 To print records from a DAS data channel enter: D REPORT “name” RECORDS=nnn COMPACT|VERBOSE Examples of reports are: D REPORT “CONC” RECORDS=35 VERBOSE D REPORT “CALDAT” RECORDS=10 D REPORT “PNUMTC” RECORDS=155 VERBOSE Automatic RS-232 reporting can be independently enabled and disabled for each Data Channel.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 All of the default Data Channels sample more than one parameter, for these channels, each parameter is printed on a separate line. There is also a compact format. If this attribute is enabled, all 3 concentration parameters are printed on one line as shown below: D 94:08:00 0200 CONC : 20.0 120.0 100.0...
Page 80
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Variables can be changed. Before changing the settings on any variables, please make sure you understand the consequences of the change. We recommend you call the factory before changing the settings on any variables.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 6 OPTIONAL HARDWARE AND SOFTWARE Optional equipment offered with the M200A includes: 1. Rack mount with slides (P/N 01469) 2. Rack mount without slides, ears only (P/N 01470) 3. Rack mount for external pump w/o tray (P/N 0099701) 4.
Adequate zero air can be supplied by connecting a Purafil/charcoal scrubber and 5 micron particulate filter (Teledyne API P/N 000369) to the zero air inlet tubing. The zero air scrubber used in conjunction with the Zero/Span Valve option provides an inexpensive source of zero air.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 6.3 Internal Zero/Span (IZS) The IZS option includes the Zero/Span Valves described above, a temperature-controlled permeation tube oven, and a rear panel mounted zero air scrubber. The IZS system is activated by the same methods as described in Table 6-1 for the zero/span valves.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 For each mode there are seven attributes that the MODE can have that control operational details of the SEQUENCE. They are: Table 6-3: IZS Sequence Attributes Attribute No. Attribute Name...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 6-5: Example of AutoCal Setup Step Action Comment Press SETUP-ACAL This button sequence will cause the AUTOCAL menu to be displayed Press PREV-NEXT Press PREV-NEXT until SEQ 2 is displayed...
Page 86
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 If an ammonia permeation tube is being used, substitute the molecular weight of NH g/mole) for that of NO The temperature is set at 50.0 C. Check SETUP-MORE-VARS and scroll to the IZS-SET variable to verify that the temperature is properly set.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 TELEDYNE API recommends that you purchase replacement permeation tubes from: VICI METRONICS 2991 Corvin Drive Santa Clara, CA 95051 USA Phone 408-737-0550 Fax 408-737-0346 The M200A uses 560 cc/min of zero air over the perm tube. Therefore you should order a permeation tube with a SPECIFIC OUTPUT of 400 to 450 ppb at .560 liter/minute.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7 CALIBRATION AND ZERO/SPAN CHECKS There are several ways to check and adjust the calibration of the M200A. These different methods are summarized in Table 7-1. In addition, all of the methods described in this section can be initiated and controlled via the RS-232 port.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-1: Types of Zero/Span Check and Calibration Section Type of Cal or Check Description Manual Z/S Check or Calibration This calibration option uses calibration gas through the sample port coming in through the sample port.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.1 Manual Zero/Span Check or Cal With Zero/Span Gas in the Sample Port The calibration of the instrument can be checked or adjusted using gas from the sample port.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-3: Enter Expected Span Gas Concentrations Procedure Step Number Action Comment Press This key sequence causes the M200A to prompt for the CAL-CONC-NOX expected NO concentration. Enter the NO span concentration value by pressing the key under each digit until the expected value is set.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.2 Manual Zero/Span Check or Calibration with Zero/Span Valves Option The Zero/Span valve option can be operated from the front panel keyboard. In the Zero/Span valve option the zero and span gas come into the valves through ports on the rear panel of the instrument.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.3 Manual Zero/Span Check with IZS Option The Internal Zero/Span (IZS) system can be operated from the front panel keyboard. When the M200A is in the SAMPLE mode, and the IZS option is installed, the CALS (Calibrate-Span) or CALZ (Calibrate-Zero) buttons will be visible.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-7: Enabling Dynamic Zero/Span StepNumber Action Comment Press Causes the M200A go to the AutoCal menu. SETUP-ACAL Press Select the sequence you want for dynamic calibration. PREV-NEXT Press SET Select the SET menu.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.6 Calibrate on NO Permeation Tube If the M200A is equipped with the IZS option, is possible to span the instrument on the permeation tube. If this feature is enabled, the instrument internal valves are set so that gas is routed through the permeation tube, then the moly converter, thus producing NO gas.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-8: Enabling CAL-ON-NO2 Step Number Action Comment Press SETUP- Causes the M200A go to the VARS(Variables) menu MORE-VARS Enter the VARS Password value is 8 1 8 password, then...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.7 Use of Zero/Span Valves or IZS with Remote Contact Closure The Zero/Span valve or IZS options can be operated using Remote Contact Closures provided on the rear panel. See Figure 2-2 for connector location and pinout. When the contacts are closed, the analyzer will switch to zero or span mode.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8 EPA Protocol Calibration If the M200A is to be used for EPA compliance monitoring, it must be calibrated in accordance with the instructions in this section. In order to insure that high quality, accurate measurements are obtained at all times, the M200A must be calibrated prior to use.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-10: Activity Matrix for Calibration Equipment and Supplies Equipment/ Frequency And Method Action If Requirements Acceptance Limits Supplies Of Measurement Are Not Met Recorder Compatible with output Check upon receipt Return equimpent to signal of analyzer;...
Devices that condition ambient air by drying and removal of pollutants are available on the commercial market such as the Teledyne API Model 701 Zero Air Generator. We recommend this type of device for generating zero air. Detailed procedures for generating zero air are in...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.2.2 Selection of NO Span Gas Standards The NIST-SRM's provide references against which all calibration gas mixtures must be compared (Section 2.0.7, Q.A. Handbook). The procedure requires the comparison of the concentration of a commercial, working calibration standard to an NIST-SRM.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.4 Gas Phase Titration (GPT) System 7.8.4.1 GPT Principle of Operation Gas Phase Titration (GPT) is recommended for calibration of the M200A. Those using a NO permeation tube should refer to TAD.
Page 105
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Select an NO standard gas that has a nominal concentration in the range of 50 to 100 ppm. Determine the exact concentration [NO] by referencing against an NIST-SRM, as discussed in Section 2.0.7 (Q.A.
Page 106
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 NOTE If t is >2 minutes or if P is <2.75 ppm-min, changes in flow conditions ) or in the reaction chamber volume (V ), or both will have to be made, and t and P will have to be re-calculated.
Page 107
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Because low flows are difficult to control and measure, it is often advantageous to set a higher total flow than needed. In this example, we will let F = 2750 cm3/min...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.5 Dynamic Multipoint Calibration Procedure The procedure for calibration of chemiluminescent NO analyzers by GPT is specified in the Federal Register. This section applies the general procedure to the specific case of the M200A.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.5.1 Zero Calibration Procedure Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero value. The following Table 7-12 details the zero calibration procedure.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-13: Expected Span Gas Concentration Procedure Step Number Action Comment Press This key sequence causes the M200A to prompt for the CAL-CONC-NOX expected NO concentration. Enter the NO span concentration value by pressing the key under each digit until the expected value is set.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Plot the analyzer NO and NO responses versus the corresponding calculated concentrations to obtain the calibration relationships. Determine the straight line of best fit (y = mx + b) determined by the method of least squares.
Page 113
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Calculate the resulting NO concentrations as follows: Equation 7-10 ORIG Where [NO] is the NO co ncentration before the GPT ozone is turned on, and [NO] is the NO ORIG aining after GPT.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.6.1 Automatic Moly Converter Efficiency Compensation The M200A can automatically compensate the NO and NO readings for the molybdenum converter efficiency. There are 2 ways to enter the converter efficiency into the instrument. The first is to enter the efficiency as a decimal fraction using the CAL-CONC-MOLY-SET menu.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.7 Calibration Frequency To ensure accurate measurements of the NO, NO , and NO concentrations, calibrate the analyzer at the time of installation, and re-calibrate it: No later than three months after the most recent calibration or performance audit which indicated analyzer calibration to be acceptable.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-16: Definition of Level 1 and Level 2 Zero and Span Checks (from Section 2.0.9 of Q.A. Handbook for Air Pollution Measurment Systems) LEVEL 1 ZERO AND SPAN CALIBRATION...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.9 Summary of Quality Assurance Checks The following items should be checked on a regularly scheduled basis to assure high quality data from the M200A. See Table 7-17 for a summary of activities; also the QA Handbook should be checked for specific procedures.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.10 ZERO and SPAN Checks A system of Level 1 and Level 2 zero/span checks (see Table 7-1) is recommended. These checks must be conducted in accordance with the specific guidance given in Subsection 9.1 of Section 2.0.9 (Q.A.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.10.2 Precision Check A periodic check is used to assess the data for precision. A one-point precision check must be carried out at least once every 2 weeks on each analyzer at an NO concentration between 0.08...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-18: NIST-SRM's Available for Traceability of Calibration and Audit Gas Standards Nominal NIST-SRM Type Size, at STP Concentration 1683a Nitric oxide in N 50 ppm 1684a Nitric oxide in N...
Page 121
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.8.12.1 NO Working Standard Traced to NIST NO Standard First, use the NIST-traceable NO standard and the GPT calibration procedure to calibrate the NO, NO , and NO responses of the analyzer. Also determine the efficiency of the converter.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.9 Calibration of Independent Ranges or Autoranging There are additional considerations when AutoRange or Independent Ranges are selected. The M200A uses one physical range of 0-2000 ppb to cover all EPA concentration ranges. The dynamic range of the internal hardware and computer software is sufficient to cover this entire range.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.10 Calibration Quality After calibration is complete, it is very important to check the QUALITY of the calibration. The calibration of the M200A involves balancing several sections of electronics and software to achieve an optimum balance of accuracy, noise, linearity and dynamic range.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 7-19: Calibration Quality Check Step Number Action Comment Scroll the TEST The SLOPE value for NO should be 1.0 ± 0.3. If the value is function menu until not in this range, check Section 9.1.6.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 7.11 References Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50, Appendix F, Measurement Principle and Calibration Procedure for the Measurement of Nitrogen Dioxide in the Atmosphere (Gas Phase Chemiluminescence), Federal Register, 41 (232): pp 52688-52692, December 1976.
Page 126
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK 7-38...
Teledy ne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8 MAINTENANCE 8.1 Maintenance Schedule NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. Table 8-1: Preventative Maintenance Schedule Item Maintenance Interval...
Teledy ne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 8-2: Preventative Maintenance Schedule Maintenance Item Interval 6 - 12 month TEST Functions intervals Zero/Span Annually or after Calibration repairs Zero/Span Checks Dail Particulate Filter Weekly as needed...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8.2 Replacing the Sample Particulate Filter The particulate filter should be inspected often for signs of plugging or contamination. It is also common for dirt particles to absorb NO , thus causing those readings to be low.
Remove the scrubber from the pump pack by loosening the 1/4" tube fittings. Unscrew the upper end cap of the scrubber and empty the contents. Replace the charcoal with TELEDYNE API P/N 00596 or equivalent. Be sure the charcoal completely fills the canister so that flow channels will not form.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 NOTE If the instrument is to be turned off for more than 8 hours, the perm tube must be removed from the IZS oven. Even at room temperature the perm tube continues to release NO .
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8.6 Replacing the Molybdenum Converter The Molybdenum Converter is located in the center of the instrument, see Figure 2-5 for location, and Figure 8-2 for the assembly. The heater, thermocouple, converter is designed to be replaced as a single unit.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8.7 Cleaning the Reaction Cell The reaction cell should be cleaned whenever troubleshooting points to it as the cause of the trouble. A dirty cell will cause excessive noise, unstable zero or span, or low response.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8.8 Pneumatic Line Inspection Particulate matter in the pneumatic lines will affect both flow rate and response time. It is important that the pneumatic system be periodically inspected and thoroughly cleaned if necessary.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8.9 Leak Check Procedure If a leak checker is not available, it is possible to leak check the instrument using the M200A's pump plus a shut-off valve. Turn off instrument power and pump power.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 8.10 Light Leak Check Procedure Scroll the TEST functions to PMT. Input zero gas. Shine a powerful flashlight or portable incandescent light at the inlet and outlet fitting, and at all the joints of the reaction cell. The PMT value should not respond to the light.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9 TROUBLESHOOTING, ADJUSTMENTS NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. This section of the manual contains information on diagnosing and repairing instrument performance problems.
Page 144
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Check the TEST functions: A. Compare the TEST functions to the factory values in Table 2-1. This will often provide important clues as to the problem. B. Pay special attention to the NO and NO slopes: C.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.1 Operation Verification-M200A Diagnostic Techniques 9.1.1 Fault Diagnosis with TEST Variables Table 9-1 indicates possible fault conditions that could cause the TEST functions to be outside the acceptable range.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-1: Test Functions Test Function Factory Set-Up Comment RANGE 500PPB This is the Range of the instrument. In standard configuration all 3 outputs have the same range. Independent range option allows different ranges for each output.
Page 147
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment OZONE FL 60 - 90 This is the Ozone flow. It is measured by a solid state flow meter, and thus is a true indication of flow.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment AZERO Check value in AZERO is the current value of the AutoZero circuit reading. Final Test Values Values are typically 0 mV -20/+150. For values outside this range,...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment IZS TEMP ± The IZS temperature is the temperature of the permeation tube 0.4° C oven. This temperature is controlled by the CPU and is adjustable in 0.1°...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.1.2 Fault Diagnosis with WARNING Messages The M200A monitors several internal values for alarm conditions. If the condition for an alarm is met, the alarm is displayed on the front panel and the warning is transmitted out the RS-232 port.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-2: Front Panel Warning Messages Message Description SYSTEM RESET Analyzer was reset/powered on. This warning occurs every time the instrument is powered up, as in a power failure. It can also occur if the RAM or EEPROM is reset.
Page 152
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-2: Front Panel Warning Messages (Continued) Message Description PMT TEMP WARNING PMT temp. out of spec. The PMT temp has its own proportional controller on the preamp (01141) board.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.1.3 Fault Diagnosis Using DIAGNOSTIC Mode Diagnostic mode can be looked at as a tool kit of diagnostics to help troubleshoot the instrument. To enter DIAG mode press SETUP-MORE-DIAG.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-3: Summary of Diagnostic Modes DIAG Mode Description SIGNAL I/O Gives access to the digital and analog inputs and outputs on the V/F board. The status or value of all of the signals can be seen.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.1.3.1 Signal I/O Diagnostic Table 9-4: Diagnostic Mode - Signal I/O Signal Control Description DSP_BROWNOUT Display brownout is used to keep the display from getting corrupted during low line voltage conditions.
Page 156
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-4: Diagnostic Mode - Signal I/O (Continued) Signal Control Description BROWNOUT_RST Brownout reset works in conjunction with DSP_BROWNOUT. When DSP_BROWNOUT is set, the CPU sends a signal to reset the display and clear the DSP_BROWNOUT.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-4: Diagnostic Mode - Signal I/O (Continued) Signal Control Description ST_POWER_OK Status Bit - Power OK Logic High = Instrument power is on Logic Low = Instrument power is off...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-4: Diagnostic Mode - Signal I/O (Continued) Signal Control Description DAC_CHAN_0 Output of NO channel in mV. DAC_CHAN_1 Output of NO channel in mV. DAC_CHAN_2 Output of NO channel in mV.
Page 159
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 The value in PMT should come up to 2000 mV ± 500 mV in less than 15 sec. If the HVPS or the span gain adjust on the preamp card has been changed without doing a FACTORY CALIBRATE the reading in step 3 may be different than 2000 mV, since the overall calibration affects ELECTRIC TEST.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.1.3.5 Analog Out Step Test The Step Test is used to test the functioning of the 4 DAC outputs on the V/F board. The test consists of stepping each analog output 0-20-40-60-80-100% of the output. If the analog outputs are set for 0-5V full scale the outputs would step 0-1-2-3-4-5 VDC.
Page 161
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 The reaction cell pressure compensation factor is equal to 1.0000 when the cell pressure is 5"- Hg-A. The compensation factor increases sample concentration with increasing cell pressure to compensate for increased quenching of the chemiluminescent reaction at higher pressures. The pressure compensation correction has been found to work adequately up to about 8000 ft (2400m) altitude.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-5: Model 200A Internal Variables Default Value Name Units Description Value Range DAS_HOLD_OFF 0-60 Time that data is not put into DAS after CAL or DIAG modes TPC_ENABLE...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-6: Test Channel Readings TEST Channel Minimum* Maximum* Description 0 mV 5000 mV PMT detector output from the preamp. This DETECTOR signal has been amplified and filtered. Since the instrument is switched and uses AutoZero, normal values can vary from –100 to 5000 mV.
Page 164
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-6: Test Channel Readings (Continued) TEST Channel Minimum* Maximum* Description IZS TEMP The IZS temperature is set to 50 C. At the setpoint, a typical reading is 3600 mV.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.1.6 Factory Calibration Procedure The Factory Cal procedure balances the PMT, preamp, and software gain factors so the instrument has optimum noise, linearity, and dynamic range. It should be used when you are unable to zero or span the instrument, when the slope and offset values are outside of the acceptable range, or when other more obvious reasons for problems have been eliminated.
Page 166
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Calculate the expected NORM PMT mV reading. For ranges up to 2000 ppb, multiply the expected span value by 2 to get the mV reading. For ranges 2001 to 20000 ppb, multiply the expected span value by .2 to get the mV reading.
Press EXIT to return to SAMPLE mode. If this procedure does not produce an instrument that will properly span, please contact your local distributor or the Teledyne API factory. Teledyne API's phone number is on the front page of this manual.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Change in span gas concentration. A. Zero air or ambient air leaking into span gas line. B. Permeation tube or cal gas tank exhaustion. Leak in NO/NO or AutoZero switching valves.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.3 No Response to Sample Confirm correct operation of analog output by performing Analog Output Step Test in Section 9.1.3.5. Confirm general operation of analyzer. A. Check for AC Power, Section 9.2.1.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.4 Negative Output 1. Mis-calibration. The 'zero' gas that was used to zero the M200A contained some NO gas - that is, it had more NO gas than that of the sample air. May also be caused by doing a zero calibration using ambient air.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.6 Unstable Span Common causes are: Leak in pneumatic system. Light leak - check the sensor module with strong light. Instrument not fully warmed up. Sample lines or sample filter dirty - clean or replace.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.7 Unstable Zero Common causes are: Leak in pneumatic system. Perform leak check, see Section 8.9. Miscalibration. See Table 7-1. Light leak - check the sensor module with strong light.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.9 Inability to Zero If the ZERO button is not illuminated when attempting to zero, that means the reading is outside of the software gain ranges allowed. In an analog instrument it would be the equivalent to the zero pot hitting the rotation stop.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.10 Non-Linear Response Common causes are: Leak in pneumatic system, see Section 8.9. High zero background - the PMT TEST function should be near 50 mV with zero gas.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.2.11 Slow Response Contaminated or dirty sample delivery pneumatics. A. Dirty/plugged sample filter or sample lines. B. Dirty reaction cell. Sample inlet line too long. Wrong materials in contact with sample - use glass, stainless steel or Teflon.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3 Subsystem Troubleshooting and Adjustments 9.3.1 Computer, Display, Keyboard The purpose of this section is to determine if the computer subsystem electronics hardware are working properly. Asessment will be made at the board level.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Figure 9-2: CPU Board Jumper Settings 9-36...
Page 179
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Like the display, the overall functioning of the CPU can be confirmed by a simple test. Locate the CPU board on the mother board by referring to Figure 2-5.
The most common requirement is the Ready to Send (RTS) signal must be at logic high (+5V to +15V) before the modem will transmit. The Teledyne API analyzer sets pin 8 (RTS) to 10 volts to enable modem transmission. Once the physical connection has been correctly established, it is necessary to establish the correct software settings in the modem.
Page 181
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 To troubleshoot a modem connection first disconnect the RS-232 cable from the Analyzer and verify (use a DVM) that you are getting a signal on Pin 2 of the RS232 port on the Analyzer. The signal will be between -5V and -15V with respect to signal ground (pin 5).
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.3 Voltage/Frequency (V/F) Board The V/F Board consists of 16 analog input channels, each software addressable, 8 digital inputs, and 24 digital outputs, each line independently addressable, and 4 independent analog output channels.
Page 184
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Use a DVM on each of the analog output channels to confirm the correct voltages. If the voltages step, but are the wrong values, the V/F board may be out of calibration. See Section 9.3.3 for information on how to calibrate the V/F board.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.3.1 ADC/DAC Calibration Procedure Due to the stability of modern electronics, this procedure should only need to be performed if a major sub-assembly is exchanged or when the display voltage does not match the input voltage or current to the V/F card.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-8: V/F Board Switch Settings DAC # Sw 1 Sw 2 Sw 3 Sw 4 Sw 5 Sw 6 Sw 7 OFF* OFF* OFF* OFF* OFF* OFF* OFF*...
Page 187
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 After the switches and jumpers are set, turn on instrument power and complete the following: A. Press SETUP-MORE-DIAG, then press ENTR. Scroll to D/A CALIBRATION, press ENTR. B. Press ADC to select the first task, which is to calibrate the A/D converter.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.3.2 Changing Output Voltage Ranges Several different output voltage ranges can be selected by switchs on the V/F board. See Figure 9-4 for the switch settings. 9.3.4 Status/Temp Board...
Page 190
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 In addition there is a thermocouple amplifier for the Moly converter. Thermistor Temperature Amplifier Adjustments If the temperature readouts are in error: Locate the Box temp thermistor on the board and place a calibrated thermometer near the thermistor.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.4.3 Status Output Lines, External Contact Closures The Status lines consist of 2 active input lines, and 12 active output lines. Additional circuits are present on the board but currently unused. Individual lines are set or cleared under CPU control depending on the assigned alarm condition.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 9-9: Power Supply Module Subassemblies Module Description Linear Power Supply The linear power supply board takes multiple voltage inputs from the Board power transformer and produces +5, +15, -15, +12 VDC outputs. The outputs are routed to two external connectors, P2 and P3.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 PSM Diagnostic Procedures The Linear Power Supply board can be tested by checking the DCPS - TEST function on the front panel. It should read 2500 mV ± 200 mV. If the value is outside this range, individual output voltages can be tested on connector P3, see Schematic in the Appendix for pinouts.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.6 Ozone Generator CAUTION - DANGER Lethal voltages present inside Ozone Generator - DO NOT defeat electrical interlock. Turn off instrument before servicing. Unplug Generator before disassembling. The ozone generator subsystem consists of a permeation drier, flowmeter, power supply - generator module, and power switch.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.7 Flow/Pressure Sensor The flow/pressure sensor board consists of 2 pressure sensors and a flow sensor. See Figure 9-8 for a diagram of this board. From these three sensors four values are computed and displayed on the front panel TEST funcions.
Page 200
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 In SAMPLE mode, scroll the TEST functions to SAMP FLW. Observe the reading. Then subtract the observed reading from the desired reading. Go to the VARS menu by SETUP-MORE-VARS, then pressing ENTR. Scroll to SFLOW_SET and press EDIT.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.8 Sensor Module Figure 9-9 and Figure 9-10 show the assembly of the NO sensor module. 9.3.8.1 PMT The PMT detects the light emitted by the reaction of NO with ozone. It has a gain of about 500,000 to 1,000,000.
Page 205
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Figure 9-11: PMT Cooler Subsystem 9-63...
Page 206
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.8.4 PMT Cooler The PMT cooler uses a Peltier cooler supplied with DC current from the switching power supply in the Power Supply Module. An overall view is shown in Figure 9-9. The temperature is controlled by a proportional temperature controller located on the Preamp board.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.9 Z/S Valves & IZS Permeation Tube Oven The Z/S Valves and IZS are both options in the instrument. Before troubleshooting this sub- assembly, check that the options were ordered, and that they are enabled in the software.
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 9.3.10 Pneumatic System The pneumatic system is diagramed in Figure 8-4, Figure 8-5, Figure 8-6, depending on which options the instrument was ordered with. 9.3.10.1 Leak Check CAUTION When doing a leak check do not pressurize the M200A to greater than 15psi.
Page 210
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK 9-68...
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 10 M200A SPARE PARTS LIST Table 10-1: Teledyne API M200A Spare Parts List Part No. Description 020420000 Teledyne API Model 200A Spare Parts List 000940100 Orifice, 3 mil, 60 cc (IZS)
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 10-1: Teledyne API M200A Spare Parts List (Continued) Part No. Description 011310000 Drier Assembly Complete with Flow Control 011320000 Assembly, Moly Converter, M200A 011330000 O3 Generator Assembly 011830000...
Page 213
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 10-1: Teledyne API M200A Spare Parts List (Continued) Part No. Description 024710000 Tubing: 6’, 1/8” CLR 024720000 Tubing: 6’, 1/8” BLK 024730000 Tubing: 6’, 1/4” BLK 024750000 Tubing: 6’, 1/4” TYGON...
Page 214
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 10-1: Teledyne API M200A Spare Parts List (Continued) Part No. Description KIT000055 IZS Retrofit, M200A (WithValves) KIT000056 IZS Retrofit, M200A (w/o Valves) KIT000067 Retrofit, M200A Rcell Rebuild Kit...
Page 215
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 10-2: Teledyne API MODEL 200A 37 mm Filter Expendables Kit Part No. Description 014020200 M200A 37 mm Filter Expendables Kit Includes: 002270100 Gasket (Rx Cell) Qty. 12 003690100 Filter, TFE, 37 mm, Qty.
Page 216
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 10-4: Teledyne API MODEL 200A Expendables Kit - IZS Part No. Description 014030000 M200A Expendables Kit - IZS Includes: 005970000 Purafil, 6 LBS, IZS OR Valve VER. FL0000001...
Page 217
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 Table 10-6: Teledyne API MODEL 200A Spares Kit for 1 Unit Part No. Description 014040100 M200A Spares Kit for 1 Unit Includes: 012360000 Fan, Power Supply Module 013140000 Fan, PMT Cooler...
Page 218
Teledyne API Model 200A NO Analyzer Instruction Manual, 02246, Rev. F1 INTENTIONALLY BLANK 10-8...