4.5.11 Analog Output
The analog output signal from the Model 2000RS is a voltage (0-1
VDC) or current (4-20 mA) signal which depends on the concentration
AND the currently activated analysis range. To relate the signal output
to the actual concentration, it is necessary to know what range the
instrument is currently on, especially when the analyzer is in the auto
ranging mode.
The Analog Output screen allows the user to choose the output
configuration for the range ID and the concentration. The available
choices are 0-1 VDC or 4-20 mA. Both outputs are not available at the
same time.